PN 3622296
April 2014 Rev. 1, 1/18
© 2014-2018 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.
E-DWT Power Supply Replacement Kit
P/N 3622351
Instruction Sheet
Power Supply Replacement kit for E-DWT,
P/N 3622351, includes:
DESCRIPTION
QTY
PART NO.
12 V power supply (MFG P/N
AB-120A125-CP6-151179)
1 4766592
E-DWT to RPM4, power
cable
1 3335955
ADAPTER PN Jack 21-25
1
4808053
Introduction
This kit is intended to provide a new power supply and
cable for E-DWT. It replaces MFG P/N
TR9CE3350CCP-Y with correct power supply MFG
P/N AB-120A125-CP6-151179.
Procedure
1. Vent the unit and disconnect the power cable.
2. Remove all external cables and plumbing
connected to the unit.
3. Plug test port, close reservoir vent, and make sure
reservoir cap is tightened (see Figure 1).
Figure 1: E-DWT top view
4. Remove one of the four variable volume handles to
allow the E-DWT to be placed on its side (see
Figure 2)
Figure 2: E-DWT on its side with one handle removed
5. Locate the 12V power supply connection on the rear
housing of the E-DWT. This connection and the
internal cable will be replaced (see Figure 3).
Figure 3: E-DWT rear view detail: power supply
connection