background image

 

PN 3622296 
April 2014 Rev. 1, 1/18 

© 2014-2018 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. 
All product names are trademarks of their respective companies. 

E-DWT Power Supply Replacement Kit 

P/N  3622351

 

Instruction Sheet

 

 

Power Supply Replacement kit for E-DWT,  
P/N 3622351, includes: 

DESCRIPTION 

QTY 

PART NO. 

12 V power supply (MFG P/N 
AB-120A125-CP6-151179) 

1 4766592 

E-DWT to RPM4, power 
cable 

1 3335955 

ADAPTER PN Jack 21-25 

4808053 

 

Introduction 

This kit is intended to provide a new power supply and 
cable for E-DWT. It replaces MFG P/N 
TR9CE3350CCP-Y with correct power supply MFG 
P/N AB-120A125-CP6-151179. 

Procedure

 

1.  Vent the unit and disconnect the power cable. 
2.  Remove all external cables and plumbing 

connected to the unit. 

3.  Plug test port, close reservoir vent, and make sure 

reservoir cap is tightened (see Figure 1). 
 

 

Figure 1:  E-DWT top view

 

4.  Remove one of the four variable volume handles to 

allow the E-DWT to be placed on its side (see 
Figure 2) 

   

Figure 2:  E-DWT on its side with one handle removed

 

5.  Locate the 12V power supply connection on the rear 

housing of the E-DWT. This connection and the 
internal cable will be replaced (see Figure 3). 

 

Figure 3:  E-DWT rear view detail: power supply 

connection

 

Отзывы: