Vital Signs Simulator
Oximeter SpO2 Optical Emitter and Detector
59
W
Warning
The pulse oximeter component of the device is not intended to
validate the SpO
2
accuracy of pulse oximeter equipment.
The pulse oximeter component of this device is not intended to
confirm the SpO
2
accuracy of the calibration curve of the pulse
oximeter monitor or to assess the optical characteristics of
representative pulse oximeter probes to determine their proper
calibration.
Not all functional testers and pulse oximeter equipment are
compatible. Functional testers can vary in pulse methods, pulse
contours, and amplitude. A functional tester might not
accurately reproduce the calibration of the pulse oximeter
equipment and can yield different results between pulse
oximeter equipment.
Note
When the Type value is set to Masimo Rainbow, more SpO
2
parameters show on the display than other types of sensors. See
Test a Masimo Rainbow SpO2
to learn about these parameters.
To set oximeter SpO
2
optical emitter and detector parameters, push
to
show the screen in Figure 44 on the display.
glh018.bmp
Figure 44. Oximeter SpO
2
Optical Emitter and Detector Screen
Connect the SpO
2
artificial finger to the SpO
2
jack on the front panel as shown in
Figure 45.