Vital Signs Simulator
Simulate Cardiac Output
45
Set the Cardiac Output Waveform
To set the cardiac output waveform:
1. In the Cardiac Output screen, use
or
to highlight the wave parameter.
2. Push
.
3. Push
or
to highlight a waveform name in the list of waveforms.
Table 13 shows the cardiac output waveforms for the Product.
Table 13. Cardiac Output Waveforms
Waveform Description
2.5 L/min
Normal waveform with accuracy of 2.5 L/min
5.0 L/min
Normal waveform with accuracy of 5.0 L/min
10.0 L/min
Normal waveform with accuracy of 10.0 L/min
Interrupted Injectable
Interrupted injection waveform
LR Shunt
Left to right ventricular shunt
Calibrated Pulse
Calibrated square wave pulse
4. Push
to set the wave and go back to the Cardiac Output screen.
Set the Baseline Temperature
To set the baseline temperature:
1. In the cardiac output screen, if not already highlighted, use
or
to
highlight the baseline temperature value.
2. Push
.
3. Push
or
to highlight
36
°
C
,
37
°
C
, or
38
°
C
.
4. Push
to set the baseline temperature and go back to the Cardiac Output
screen.
Set Injectate Temperature
To set the injectate temperature:
1. In the cardiac output screen, use
or
to highlight the injectate
temperature value.
2. Push
.
3. Push
or
to highlight
0
°
C
or
24
°
C
.
4. Push
to set the temperature and go back to the Cardiac Output screen.
As you change the injectate temperature, the calibration coefficient necessary for
the monitor shows on the display.
Start a Cardiac Output Simulation
After you set the parameters for a cardiac output simulation, push the
Start
softkey. The simulation completes automatically. To stop the simulation, push the
Stop
softkey.