ProSim 8
Users Manual
10
Pre-Defined Simulations
The pre-defined simulations are set through the softkeys along the bottom of the
Home screen. See Figure 2. There are seven factory pre-defined simulations:
normal, hypertensive, hypotensive, tachycardic, bradycardic, heart attack, and
asystole. A
More
softkey shows on the display for more simulations. Table 7 lists
the parameter values for each pre-defined simulation.
Table 7. Pre-Defined Patient Simulations
Simulation Name
Parameter Pre-Set
Value
Normal
Wave Form
NSR (Adult)
ECG Rate
60 BPM
Respiration Rate
20 brpm
Temperature 37.0
°
C
IBP Channel 1
120/80 mmHg (Art)
IBP Channel 2
28/15 mmHg (PA)
NIBP
120/80 (93) mmHg
SpO
2
97
%
Hypertensive
Wave Form
NSR (Adult)
ECG Rate
130 BPM
Respiration Rate
40 brpm
Temperature 37.0
°
C
IBP Channel 1
200/150 mmHg (Art)
IBP Channel 2
45/25 mmHg (PA)
NIBP 200/150
(166)
mmHg
SpO
2
94
%
Hypotensive
Wave Form
NSR (Adult)
ECG Rate
40 BPM
Respiration Rate
15 brpm
Temperature 37.0
°
C
IBP Channel 1
60/30 mmHg (Art)
IBP Channel 2
15/10 mmHg (PA)
NIBP
60/30 (40) mmHg
SpO
2
95
%
Tachycardia
Wave Form
NSR (Adult)
ECG Rate
180 BPM
Respiration Rate
50 brpm
Temperature 40.0
°
C
IBP Channel 1
80/50 mmHg (Art)
IBP Channel 2
25/10 mmHg (PA)
NIBP
80/50 (60) mmHg
SpO
2
88
%