5.1
TERMINAL CONNECTORS INTRINSICALLY SAFE APPLICATIONS
The unit is classified as group IIB/IIIC by default.
Classification of the unit as group IIC is only possible under the following conditions:
The indicator is either supplied by
the internal supply (option -PC);
the external supply connected to terminals 0 and 1 (option -PD);
The maximum values for any of those circuits are those as defined for group IIB/IIIC;
No other active external Intrinsically safe circuits may be connected to the indicator, with
exception of circuits connected to terminals 3 and 4 and/or terminals 5 and 6; the maximum
values for any of those circuits are those as defined for group IIB/IIIC.
Terminal connectors F130-P-XI:
For intrinsically safe applications, consult the safety values in the certificate.
Fig. 25: Overview terminal connectors XI
– Intrinsically safe applications