Page 42
FW-F030-P-M_v0404_01_EN.docx
5.4.3
F030-P-OT-PX-XI-ZB - Ex ia IIC/IIIC
Fig. 39: F030-P-OT-PX-XI-ZB - Ex ia IIC/IIIC - Intrinsically Safe application.
*
Sensor supply voltage for pulse type P: Terminal 3: 1.2-3.2V DC.
Please note: Type PX may be used in combination with the battery (Type PC).
PX will power the F030-P, the battery will be disabled automatically until the power is disconnected.
5.4.4
F030-P-OT-PD-XI-ZB - Ex ia IIC/IIIC
Fig. 40: F030-P-OT-PD-XI-ZB - EX ia IIC/IIIC - Intrinsically Safe application.
*
Sensor supply voltage for pulse type P: Terminal 3: 1.2-3.2V DC; Terminal 6: 8.2V DC.
Please note: Type PD may be used in combination with the battery (Type PC).
PD will power the F030-P, the battery will be disabled automatically until the power is disconnected.
Configuration example IIA - IIB and IIC application - F030-P-OT-PX-XI-ZB
I.S. flowmeter
input type: P
pulse
TERMINAL CONNECTORS
F0-series
1
2
3
Common ground
Signal
Supply *
C
irc
ui
t d
ep
en
ds
o
n
ty
pe
o
f s
ig
na
l
7
8
Common ground
4
5
Common ground
Main supply
9
1
0
Common ground
Supply backlight
HAZARDOUS AREA
SAFE AREA
+
-
Power supply
For example
MTL5025
+
-
Power supply or
switch interface
For example
MTL5025
MTL5011B
e.g. relay
+
-
Power supply
For example
MTL5025
= max. 30 V
= max. 200 mA
= max. 1,2 W
Uo
Io
Po
= max. 30 V
= max. 200 mA
= max. 1,2 W
Uo
Io
Po
= max. 30 V
= max. 200 mA
= max. 0,75 W
Uo
Io
Po
Note: above values are safety values.
Consult the technical specification for operational values.
+
-
Power supply
For example
MTL5025
Uo
Io
Po
Power supply type PX: 8-30V DC
Backlight option: type ZB
*Sensor supply voltage for pulse type P: Terminal 3: 1.2V - 3.2V DC.
Please note: type PX may be used in combination with the battery (type PC). PX will power the unit; the battery will be disabled automatically till power is disconnected.
= max. 30 V
= max. 150 mA
= max. 0,92 W
Intrinsically Safe apparatus
Control output type OT: passive transistor
Configuration example IIA - IIB and IIC application - F030-P-OT-PD-XI-ZB
TERMINAL CONNECTORS
F0-series
1
2
3
Common ground
Signal
Supply *
C
irc
ui
t d
ep
en
ds
o
n
ty
pe
o
f s
ig
na
l
7
8
Common ground
6
4
5
Common ground
Main supply
Supply *
9
1
0
Common ground
Supply backlight
HAZARDOUS AREA
SAFE AREA
+
-
Power supply
For example
MTL5025
+
-
Power supply or
switch interface
For example
MTL5025
MTL5011B
e.g. relay
+
-
Power supply
For example
MTL5025
= max. 30 V
= max. 200 mA
= max. 1,2 W
Uo
Io
Po
= max. 30 V
= max. 200 mA
= max. 1,2 W
Uo
Io
Po
= max. 30 V
= max. 200 mA
= max. 0,75 W
Uo
Io
Po
Note: above values are safety values.
Consult the technical specification for operational values.
Power supply type PD: 16-30V DC
Backlight option: type ZB
*Sensor supply voltage for pulse type P: Terminal 3: 1.2V - 3.2V DC. Terminal 6: 8.2V DC.
Please note: type PD may be used in combination with the battery (type PC). PD will power the unit; the battery will be disabled automatically till power is disconnected.
Intrinsically Safe apparatus
Control output type OT: passive transistor
I.S. flowmeter
input type: P
pulse
Содержание F030-A
Страница 53: ...Page 53 FW F030 P M_v0404_01_EN docx...
Страница 54: ...Page 54 FW F030 P M_v0404_01_EN docx...
Страница 55: ...Page 55 FW F030 P M_v0404_01_EN docx...