background image

 

 

 

 

 

 

 
C300.001TC, C300.10TC 
Stainless Steel Conductivity Sensor 
 

INSTRUCTION MANUAL 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

            

EN 09-10 

 

 

1. General data 

These special sensors are particularly suitable for conductivity measurement in 
applications where stainless steel is requested as material in contact with fluid. 
Sensors are completely sealed and consist on: 

      _ a double coaxial electrodes conductivity cell, 
      _ a Pt1000 temperature sensor with 3 wires connections, 
      _ cable connection wires are marked by different colours.   

 

 

 

2. 

Technical features

 

Wetted Materials         

  

 :  

Sensor Body: Stainless Steel AISI 316L 
O-ring     

  

    : EPDM 

 

Temperature@Pressure: 100°C @ 7 bars (installed in PP fitting);  
                                     

  

  100°C @ 34 bars (installed in SS fitting); 

                                    

  

   120°C @ 13 bars (installed in SS fitting). 

 
Cable               

   

        

  

  : 3 meters of shielded cable with 5 wires 

 
Cell constant: 0,01 ; 10,0 
 
 

3. Installation 

Conductivity sensor should be installed properly to gain best performances. 
It should obey these rules: 
_ sample in contact with sensor should be representative of fluid to measure; 
_ liquid should continuously flow throw the cell; 
_ flow rate should avoid cavitation phenomenas; 
_ cell mounting should be so that an air bubbles presence is avoided on electrodes 
surface; 
_ solid sediments shouldn’t accumulate into space between electrodes dedicated to 
measurement; 
_ it should verify that temperature and pressure conditions are obeyed. 
 
 
 
 
 