6.34. Dose Rate Alarm Thresholds
FLIR Detection
Settings and Commands
Alarm
The value with unit for the alarm threshold. Use “Off” (the lowest value) to suppress this
alarm. Available units are
nSv – µSv – mSv – Sv (nrem – µrem – mrem – rem)
Warning
The value with unit for the warning threshold. Use “Off” (the lowest value) to suppress this
warning. Available units are
nSv – µSv – mSv – Sv (nrem – µrem – mrem – rem)
Down
,
Select
Choose and select menu items, see
2.2.2
, p.
35
for details.
+
,
-
,
Accept
,
Cancel
Edit values, see
2.2.3
, p.
37
for details.
Left
,
Right
Edit composite values, see
2.2.4
, p.
40
for details.
Exit
Return to the superior menu.
6.34
Dose Rate Alarm Thresholds
Options Menu (p.
ä
Alarm Options (p.
)
ä
Dose Rate Alarm Thresholds
Settings (p.
ä
Alarm (p.
ä
Gamma Alarms (p.
189
)
This screen (Figure
176
, p.
163
) groups the thresholds for dose rate alarms and warnings.
Specify the thresholds by setting a value with a unit. The base unit (Sv/h or rem/h) available here
re lects your preferences (see
6.31
, p.
159
).
You can specify values between 1 nSv/h and 1 Sv/h (100 nrem/h to 100 rem/h).
For meaningful alerts, specify an alarm threshold above the warning threshold.
Settings and Commands
Alarm
The value with unit for the alarm threshold. Use “Off” (the lowest value) to suppress this
alarm. Available units are
nSv/h – µSv/h – mSv/h – Sv/h (nrem/h – µrem/h – mrem/h – rem/h)
162
identiFINDER
®
R300/en/2014.4(13623)/Feb2015