4226044 (Rev. C) © 2020 FLIR Systems, Inc.
31
FLIR Fido® X4 Operations Manual
5.1.2.1 Startup Wizard
While the Fido X4 is heating, you may choose to play the Startup Wizard. The Startup Wizard
is a series of short videos, without sound, which show the procedure for verification,
sampling, and changing the sensing element. Press
during startup to begin the Startup
Wizard.
5.2 Verifying the System
After the sensor has reached operating temperature and before operation begins, the Fido
X4 requires verification. The system will display the message “Present Reference Stand-
ard.”
Note:
For more information on the reference standard, see 3.5.3 on page 20.
Figure 5-3
: Present Reference Standard screen
You can postpone the verification process for a short time by selecting
LATER
. However,
the system will display a message which describes the effects of delaying verification.
Figure 5-4
: Incomplete verification warning
Follow the on-screen prompts or these instructions to perform the verification process.