Operation — Operator instructions (Auto and Button mode)
4. The camera must first record an initial sample for the reference temperature
calculation.
Place a healthy person at the marking on the tape that indicates the screening posi-
tion. The camera graphics may indicate that the person needs to move closer to the
camera, which is OK to do.
5. Make sure the person looks into the camera and stands still until a screening result is
displayed.
6. Check that the camera focus is correct and that the measurement was done at the
tear duct. If not, deactivate and activate the Screening mode again and restart the
initiation.
7. Wait for two minutes. Make sure there is no people or movement in front of the cam-
era. During this time, the camera adapts to the ambient temperature.
8. Place a healthy person at the marking for the screening position (not closer or further
away from the camera).
9. Check that:
• A screening result is displayed with the person standing at the screening position
(not closer).
• The measurement was done at the tear duct. If not, deactivate and activate the
Screening mode again and restart the initiation.
• The camera focus is correct. If not, adjust the focus by rotating the focus ring or by
pushing the autofocus button.
• The camera is now ready to screen visitors.
Note
If the camera is turned off, the system must be initiated again.
12.3.2
Screening procedure
Once the screening station is up and running, the screening station normally does not re-
quire any action from you. Depending on your workflow, you may need to be available for
guidance and alarms.
This is what happens during the screening:
1. The visitors enter the queue to the screening station.
2. One visitor at the time goes to the screening position.
Note
• Removing eyeglasses is crucial for a correct screening result.
• It is important to stand at the screening position (not closer or further away from
the camera).
• The eyes must not be covered.