7. Aim the camera toward a point of interest. The object should be within the frame of
the measurement box.
8. To reset the sampled average, push and hold the programmable button
.
9. To sample, push the programmable button
.
10. Aim the camera toward more points of interest. Sample 10 times to build up a sample
base by pushing the programmable button
.
The alarm is now set up and ready to use. Occasionally record a few samples if the
alarm is used for a long time or if the conditions change.
Note
• Every time you push the programmable button
, a sample is saved. Make sure you
aim the camera toward an object within the interesting temperature range when you
push (and hold) the button.
• The algorithm has a memory of the last 10 samples. It discriminates between the high-
est and lowest values, and calculates an average of the remaining values.
• Do not modify the measurement setup or activate another alarm because this will de-
activate the screening alarm.
#T559954; r. AT/65353/65353; en-US
100
Содержание 72501-0101
Страница 2: ......
Страница 3: ......
Страница 5: ...User s manual FLIR T10xx series T559954 r AT 65353 65353 en US v ...
Страница 6: ......
Страница 14: ......
Страница 43: ...Screen elements 11 T559954 r AT 65353 65353 en US 29 ...
Страница 125: ...Mechanical drawings 30 See next page T559954 r AT 65353 65353 en US 111 ...
Страница 134: ...Digital I O connection diagram 32 See next page T559954 r AT 65353 65353 en US 120 ...
Страница 135: ......
Страница 136: ...CE Declaration of conformity 33 See next page T559954 r AT 65353 65353 en US 122 ...
Страница 137: ......
Страница 144: ... T559954 r AT 65353 65353 en US 130 ...
Страница 145: ......