Junction Box
Type KAK-4
T
E
C
H
N
I
A
L
M
U
Flintec GmbH Bemannsbruch 9 74909 Meckesheim GERMANY
www.flintec.com
Страница 1: ...Junction Box Type KAK 4 T TE EC CH HN NI IC CA AL L M MA AN NU UA AL L Flintec GmbH Bemannsbruch 9 74909 Meckesheim GERMANY www flintec com...
Страница 2: ...INTEC shall not be liable against any damages or problems arising from the use of any options or any consumable products other than those designated as Original FLINTEC Products NOTICE The contents of...
Страница 3: ...correspond to the corners of the scale i e Corner 1 Load cell 1 Corner 2 Load cell 2 etc LOAD CELL CABLE CONNECTION First the cable gland see in fig 1 must be loosened Then you have to feed the load c...
Страница 4: ...to this basic corner 5 Calculate the correction resistance as follows Deviation in kg Correction resistance Test load in kg X Input resistance of the load cell 1100 input resistance for BK2 SB4 SB5 S...