FT-112(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
40
of
170
Format Alibi Memory:
Warning: Alibi memory formatting should be done by authorized person only. Otherwise erasing alibi memory
may bring you undesired legal responsibility.
You may need to format alibi memory after installing the used SD card, to erase previous records before
starting usage in legal or to erase alibi memory if you don’t need previous records legally. This process runs if
the calibration switch is at programming position.
1. Go to the parameter [ 8-- METROLOGY ] main block in set up.
2. Press key 7 times to access parameter 816.
3. Select YES and press
key.
4. The warning prompt [ ARE YOU SURE? ] appears on the display.
5. Press
key to start formatting or Press
key to escape.
6. Press
key 3 times to return operation.
The empty fields are indicated as [ ------- ] and transferred as [ NO RECORD ]; and the corrupted records
are indicated as [ Error ] and transferred as [ CORRUPTED ].