FT-112(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
35
of
170
Advanced Functions
Programmable keys
,
and
keys can be programed to fit the instrument to your application and for easy use.
The programmable functions are: high resolution, temporary gross indication, unit change etc. Refer to sub-
block 24-,
High resolution
If you press
key, the weight value is displayed 10 times higher. Press the same key again to end. High
resolution is displayed temporarily for 5 seconds at approved scales. High resolution cannot be printed.
Unit change
Pressing the
key, you can change between two units. The unit change is available between kg and lb.
Refer to parameters 521.
Dynamic weighing
APPLICATION: Living stock weighing, weighing of unstable load, unstable industrial weighing systems.
RELATED PARAMETERS: Sub-block 33-, page 59.
The dynamic weighing can be programmed for single weighing or continuous weighing after assigning one of
the programmable keys for dynamic weighing. Single weighing is used for dynamic weighing like living
animals; continuous dynamic weighing can be used for loads which have very high vibration like reactors with
high speed mixers, conveyors etc.
Single weighing operation:
1. Place container on the scale.
2. Press
key to tare the scale.
3. Load the scale.
4. Press the dynamic weighing key to start dynamic weighing. Or dynamic weighing is started
automatically if the load is heavier than 50 divisions. Refer to parameter 331.
5. After the weighing cycle, the dynamic weight value is displayed.
6. Unload the scale or press the dynamic weighing key to reset dynamic weighing operation.
Depends on the operation type selected at parameter 331, the dynamic weighing may start automatically if the
load is heavier than minimum weight or may end after unloading the scale (refer to
page 59
).
Specified tare PT (Preset tare)
If you have specified containers, you may enter their weight values numerically using preset tare (PT) for
taring the scale. This feature eliminates the taring of empty container to measure the net value in it, to reduce
your operation time by weighing only filled containers to get net weight value. To be able to use this feature, it
must be enabled first. Refer to
Preset Taring with numeric entry:
1. Press
key.
2. Enter tare value numerically.
3. Press
key for taring or press
key to escape.