FT-111(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
93
of
146
BSI-Base Commands
A
Read all weight data
Command
: [ADR][A]
Response
: [ADR][A][STATUS][SIGN][NET W][SIGN][TARE W][SIGN][GROSS W]
Example :
Command
: 01A
Response
: 01AS+000123.4+000111.1+000234.5
01AD+000123.4+000111.1+000234.5
01AO ( ADC out error )
Comments
:
The response is net, tare and gross weight values or error status.
All weight data is transmitted immediately after receiving the command.
B
Read Gross weight
Command
: [ADR][B]
Response
: [ADR][B][STATUS][SIGN][WEIGHT VALUE]
Example :
Command
: 01B
Response
: 01BS+000123.4 (gross weight is stable and 123.4)
01BD+000123.4 (gross weight is dynamic and 123.4)
01B−
(under load)
Comments
:
The response is the gross weight value (stable or dynamic) or error status.
Gross weight data is transmitted immediately after receiving command.
C
Clear the tare memory
Command
: [ADR][C]
Response
: [ADR][C][A]
(Cleared and the scale is in gross mode)
Comments
:
The response status is always Ack in weighing or force mode.
G
Read voltage value of DC power supply
Command
: [ADR][G]
Response
: [ADR][G][STATUS][VOLTAGE VALUE]
Example :
Command
: 01G
Response
: 01GA234
(Power supply is 23.4 VDC)
01GA150
(Power supply is 15.0 VDC)
Comments
:
Voltage value is 3 byte and sends with 0.1 V increment.
I
Read indicated weight
Command
: [ADR][I]
Response
: [ADR][I][STATUS][SIGN][WEIGHT VALUE]
Example :
Command
: 01I
Response
: 01IS+000123.4 (Weight is stable and 123.4)
01ID+000123.4 (Weight is dynamic and 123.4)
01I+
(Overload)
Comments
:
Indicated weight value (stable or dynamic) is transmitted immediately.
The weight value may be in gross or net.
P
Print :Read the stable weight
Command
: [ADR][P]
Response
: [ADR][P][STATUS][SIGN][WEIGHT VALUE]
Example :
Command
: 01P
Response
: 01PS+000123.4 (Weight is stable and 123.4) or
01PN
(Could not print)
Comments
:
Checks status and it must be stable. Else Nack status is send.
Status can be Stable or Nack.