FT-1x Quick Start, May 2012

 

FT-1x  Quick Start 

 

Important hint: 
This instruction note does NOT replace the technical manual 
 It is only a brief survey for the commissioning of a type FT-1x 
indicator. 

 
 
 

1.

 

All mechanical assembly has to be done properly 

2.

 

Execute electrical connections 

 

Connect load cell with FT-1x: 

 

 

 

 

 

 
 
 
 

 

Optional FT-1x to PC communication: 

RS232,  

PC typical allocation 

Definition 

D-Sub, 9-pole, female 

Desk type housing and  

panel type housing 

Plug connector J11 

Stainless steel housing 

TxD 

RxD 

GND 

Shield 

Connector casing 

 

 

3.

 

Setup the scale 

 

1.

 

Turn on the instrument and wait for the start up. 

2.

 

Access the menu for setup and calibration. Here you have 
to press following keys in sequence: 

 
<

F>

  and  <

Info>

  and  

->0<-

  and  <

Info>

  and  <

Enter>

 

 
 

With the 

<↑>

 and <

↓>

 keys you select the parameter block 

[XXX]

, with the <

Enter>

 key you confirm 

your selection. The <

F>

 key is used to exit the setup. 

 

3.1

 

Weight display

 

 

Select parameter 

[212]

 and confirm with the <

Tare

> key. 

[CAP]

 

The capacity of the scale will be entered by the <

Tare

> and <

Zero

> keys. 

The value will be confirmed by pressing the <

Enter

> key. 

[d]

 

The display resolution of the scale will be selected by the <

Zero

> key. 

The selection will be confirmed by pressing the <

Enter

> key. 

 

6-wire

 load cell 

4-wire

 load cell 

D-Sub, 9-pole, female 

Desk type housing and  

panel type housing

 

Plug connector J12 

Stainless steel housing

 

+ Excitation 

+ Excitation 

+ Sense 

+ Excitation 

Shield 

Shield 

- Sense 

– Excitation 

– Excitation 

– Excitation 

+ Signal 

+ Signal 

– Signal 

– Signal 

Shield 

Shield 

Connector casing 