Operator's Manual
|
7-1
7 Ventilator Quick Check Procedure
7.1
Introduction
Upon initial setup of the ventilator, verify proper ventilator operation by
performing the Quick Check Procedure.
This procedure is intended to assist qualified operators to establish a routine
program for verifying proper FLIGHT 50 Ventilator operation. Perform this
procedure each time the ventilator is prepared for clinical use.
Repeat the Quick Check Procedure each time the ventilator is placed on a
new patient or the patient circuit/exhalation valve is changed.
Before performing the test, you must perform a pretest inspection, and set
up the ventilator for the test.
The Quick Check Procedure includes the following tests:
Checking the power management
Checking the alarms
Checking the monitored parameters
HOME CAREGIVERS: This procedure should be performed by
your Homecare equipment provider, prior to delivery of the
FLIGHT 50 Ventilator, to verify proper operation. It can also be
performed in the homecare environment to ensure proper setup
and function of the ventilator.
Do not use the FLIGHT 50 Ventilator if it fails this
procedure.
7.1.1 Setting Up the Ventilator for the Test
Before performing the test, do the following:
Remove the three screws from the Filter Cover. Inspect the filter.
Replace the filter if it is dirty. Reinstall the screws.
Examine the 500 ml test lung and the patient circuit to ensure that
there are no holes that will cause leaks.
Verify that the AC power cord does not have frays or breaks.
To set up the ventilator for the test:
1. Connect the detachable and integral batteries.
2. Connect the AC power cord to an AC power source.