Operator's Manual
|
6-13
6.1.1 FLIGHT 50 Ventilator
Wipe clean the FLIGHT 50 Ventilator between patients, and once a week
while in use.
To clean the ventilator:
1. Wipe clean the exterior (besides the screen) of the ventilator and all
parts not in direct contact with patients, using a cloth that has been
dampened with a medical detergent or alcohol-based cleaning solution.
2. Clean the front panel display (the screen) using a lint free damp cloth
dampened with LCD cleaner solution.
3. Air dry.
Do not apply the cleaning solution directly on the screen.
On the front panel display or ventilator housing, do not
use agents that contain acetone, toluene, halogenated
hydrocarbons, or strong alkaline.
Never autoclave or EtO sterilize the FLIGHT 50 Ventilator
and its accessories. These processes will damage the
FLIGHT 50 Ventilator and accessories, rendering them
unusable.
6.1.2 FLIGHT 50 Ventilator Accessories
All accessories should be thoroughly cleaned, rinsed, and air dried prior to
disinfecting. Examine all accessories for excessive wear or damage. Discard
and replace if necessary.
6.1.3 Humidifier Assembly (V51-00000-60 only)
Wipe clean and disinfect the Humidifier once a week while in use and clean
and sterilize between patients.
The following instructions are for the V51-00000-60 integrated
humidifier. If a third party humidifier is used see humidifier
manufacturer’s instructions for cleaning. See Figure 54 for
humidifier assembly.