Operator's Manual
|
5-3
20.
To power down the FLIGHT 50 Ventilator after it is removed from the
patient, press the
On/Standby
button twice. An audible beep sounds and
the FLIGHT 50 Ventilator automatically shuts down. Press the
Silence/Reset button to mute the audible alarm.
5.5
Built-In Humidifier
You will need to use an alternate source of humidification with
FLIGHT 50 Ventilator.
HOME CAREGIVERS: The patient’s physician or your
Homecare Dealer will determine the appropriate type of
humidification device to be used. Proper training is required
prior to setting up and using the FLIGHT 50 built-in humidifier.
The FLIGHT 50 humidifier operates only on A.C. power. It
turns off automatically when A.C. power is disconnected,
when the humidifier bottle is opened to refill the water and
after sixty (60) minutes in the Setting condition.
The FLIGHT 50 Ventilator built-in humidifier can only be
powered on by the user. It never powers on automatically.
When the FLIGHT 50 Ventilator is powered by internal battery,
appropriate actions, such as using an HME, should be taken to
prevent significant humidity deficit in the patient.
When a humidifier malfunction is detected, or a humidifier
alarm exists, the humidifier turns OFF. The user is alerted
by both an audible alarm and a change in the LED from
green to blinking yellow. The nature of the alarm is
displayed in the message window. Pressing the
Silence/Reset
button silences the audible alarm and cancels
the latched alarm indicator. Press the Humidifier On button
to restart the humidifier after the alarm condition is
corrected.
The humidifier’s round heating mechanism gets hot when
the humidifier is on. Do not touch!