Operator's Manual
|
5-2
With built-in humidifier includes (all items listed above plus):
1 ea.
V12-32000-18
Humidifier Bottle – Upper
1 ea.
V12-31001-67
Humidifier Bottle – Lower
1 ea.
V12-33001-03
Heat Sink
1 ea.
V12-60000-69
Dual Airway Temperature Probe
1 ea.
V12-70000-27
Flex Hose – Humidifier Connection
1 ea.
V12-34001-62
Absorbent Paper (pk. of 10
papers)
The built-in humidifier on the FLIGHT 50 Ventilator only
functions when used on A.C. power. It provides temperature
monitoring and alarm functions. When using the FLIGHT 50
Ventilator on battery, for transport or in home care
environments for example, alternate humidification devices
may be necessary.
For a complete list of FLIGHT 50 Ventilator accessories, see
Appendix B.
5.3
Ventilator Set Up Procedure
Figure 47 – Power Cord Restraint
To set up the ventilator:
1. Mount the ventilator on a stable surface (e.g., bedside table or the
Compact Stand Assembly).
2. For Compact Stand Assembly, follow the instructions provided with stand
to position the ventilator on pedestal mount and secure using the screws
provided.
3. Remove the white plastic cord restraint next to the power entry module.
4. Slip the power cord into the restraint.
5. Plug the power cord into the power entry module and re-attach the
restraint with cord onto the side of the FLIGHT 50 Ventilator.