Operator's Manual
|
3-16
The Battery Empty Alarm and the Device Alert Alarm cannot
be silenced permanently. These alarms indicate that an
alternate source of ventilation must be utilized. See pages
3-23 and 3-25 for more details.
3.2.21 Paw Meter (airway pressure meter)
Figure 22 – Paw Meter
Range:
–10 to 100 cmH
2
O / 98 mbar
The Paw meter displays airway pressure. It also indicates the internal
battery charge level when the Int. Battery button is tapped.
3.3
Front Panel Alarms
The front panel alarm LED indicators blink when an alarm limit setting is
violated. Once the violation is no longer in effect, the indicators latch
(remain steadily lit) until they are reset by pressing the
Silence/Reset
button.
3.3.1 High
Paw and Low
Paw Alarm Control and Display (Airway
Pressure)
Figure 23 – High Paw and Low Paw Alarm Control and Display
Range:
Paw –10 to 100 cmH
2
O / 98 mbar
High Paw Alarm 4 to 99 cmH
2
O / 97 mbar
Low Paw Alarm 3 to 98 cmH
2
O / 96 mbar
The Paw button allows the selection of the high (peak) airway pressure
alarm setting.
The Paw button allows the selection of the low peak airway pressure
alarm setting.
To adjust either alarm:
1. Press the desired button once.