16 Outputs
110
16
Outputs
If the transmitter is equipped with outputs, they have to be installed and activated before they can be used:
• assignment of a measuring channel (source channel) to the output (if the transmitter has more than one measuring channel)
• assignment of the physical quantity (source item) to be transmitted to the output by the source channel, and the proper-
ties of the signal
• definition of the output behavior in case no valid measured values are available
• activation of the installed output in the program branch
Output Options
16.1
Installation of an output
All outputs are installed in
Special Funct.\SYSTEM settings
\
Proc. outputs
.
The source items and their scroll lists are described in Tab. 16.1.
Note!
The settings will be stored at the end of the dialog. If the menu item is quit by pressing key BRK, the
changes will not be stored.
Select
Special Funct.\SYSTEM settings\Proc. outputs
.
Press ENTER.
Select the output to be installed. Press ENTER.
The scroll list contains all actually available outputs. A tick (
) after a list item indicates that
this output has already been installed.
Select
active
if all current outputs to be installed, are to be operated actively.
Select
passive
if all current outputs to be installed, are to be operated passively.
This display will only be indicated if the transmitter is equipped with a switchable current
output.
Select
yes
to install or reconfigure the output. Press ENTER.
Select
no
to uninstall the output and to return to the previous menu item in order to select
another output. Press ENTER.
Select the measuring channel to be assigned to the output as the source channel in the
scroll list. Press ENTER.
This display will not be indicated if the transmitter has only one measuring channel.
Select the physical quantity (source item) to be transmitted to the output by the source
channel.
If a binary output is configured, only the list items
Limit
and
Impuls
will be displayed.
SYSTEM settings
↕
Proc. outputs
Install Output
↕
Current I1
(
)
Loop I1,I2
passive
I1 enable
no
>YES<
I1 Source chan.
↕
Channel A:
Source item
↕
Measuring value
Содержание FLUXUS WW
Страница 1: ...User manual UMFLUXUS_WWV1 1EN Ultrasonic flowmeter for water and wastewater FLUXUS WW...
Страница 4: ...FLUXUS WW UMFLUXUS_WWV1 1EN 2018 02 28...
Страница 8: ...FLUXUS WW Table of contents 8 UMFLUXUS_WWV1 1EN 2018 02 28...
Страница 144: ...FLUXUS WW D Conformity declarations 144 UMFLUXUS_WWV1 1EN 2018 02 28 D Conformity declarations...
Страница 145: ......
Страница 146: ......