14 Outputs
14.1 Configuration of a binary output
FLUXUS WD
2018-07-16, UMFLUXUS_WDV1-3EN
112
14
Outputs
In addition to the analog outputs, the transmitter can be
equipped with binary outputs.
A binary output switches if one of the following switching
condition is met:
• the measured value exceeds or falls below the limit
• the measured value lays within or outside a defined range
• a measurement is not possible
• an event occurs
14.1
Configuration of a binary output
• Select the program branch
Options
.
• Press ENTER.
Selection of the measuring channel
• Select the channel (here:
Channel A
)
• Press ENTER.
This display will not be indicated if the transmitter has only one measuring channel.
Assignment of an output
• Select
Outputs
.
• Press ENTER.
• Select the output to be assigned to the channel.
• Press ENTER.
If the channel has already been assigned to a channel, it is displayed as follows:
Binary B1
(A)
.
• Select
Yes
to change the settings for an already assigned output or to assign a new output.
• Select
No
to cancel the assignment and to return to the previous menu item.
• Press ENTER.
Assignment of a source item
Depending on the selected source item, status or event values can be output, see Tab. 14.1.
Options
Channel x
Outputs
Select the output
Enable the output
Source item
Options\Channel A
Options\Channel A\Outputs\Binary B1(--)
Options\Channel A\Outputs\Binary B1\Enable B1
Options\Channel A\Outputs\...\Source item
Содержание FLUXUS WD
Страница 1: ...FLUXUS WD UMFLUXUS_WDV1 3EN Operating instruction...
Страница 6: ...Table of contents FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 6...
Страница 150: ...18 Settings 18 9 Key lock FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 150...
Страница 171: ...Annex FLUXUS WD D Conformity declarations 171 UMFLUXUS_WDV1 3EN 2018 07 16 D Conformity declarations...
Страница 172: ...Annex D Conformity declarations FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 172...
Страница 173: ......
Страница 174: ......