9 Measurement
FLUXUS G831ST-HT
9.6 Repeated display of measured values
93
UMFLUXUS_G831ST-HTV1-0EN, 2021-06-01
9.6
Repeated display of measured values
• Select the program branch
Measurement
to return to the measured value display.
• Press ENTER.
The following display appears:
• Select
Show measurement
in the scroll list.
• Press ENTER.
The measured values are displayed.
9.7
Execution of special functions
During a measurement, the keyboard can be used to carry out special functions.
• Press CLR. The menu item
Execute command
is displayed.
Totalizers
• Select the list item
Totalizers
.
• Press ENTER.
The following scroll list appears:
Fig. 9.6:
Scroll list in the program branch
Measurement
Measurement\Show measurement
Tab. 9.5:
Special functions
key
function
BRK
display of the scroll list in the program branch
Measurement
ENTER
display of the diagnostic window
CLR
display of the menu item
Execute command
Measurement\Execute command\Totalizers
display
description
Reset totalizers
reset the totalizer to zero
Freeze display
display the measured value of the totalizer for several seconds
Fehler zurücksetzen
reset the totalizer error
Stop/clear totalizers
stop and reset the totalizer to zero
Start totalizers
start totalizer