18 Settings
18.2 Measurement settings
FLUXUS G831ST-HT
2021-06-01, UMFLUXUS_G831ST-HTV1-0EN
134
18.2
Measurement settings
• Select the menu item
Measurement
in the program branch
Special functions
.
• Press ENTER.
• Select the menu item
Measurement settings
.
• Press ENTER.
Multi-point calibration
A multi-point calibration allows a very precise output of measuring results. It is based on calibration curves of series of
measured values.
• Select the menu item
Multi-point calibration
.
• Select
On
to activate the multi-point calibration. Select
Off
to deactivate it.
• Press ENTER.
• Select
Default
(
Multi-point calibration
=
Off
) if no customized inputs are to be made.
• Press ENTER.
If
On
is selected, a series of measured values has to be entered in the program branch
Options
.
Swift damping
If
Swift damping
is activated, each displayed measured value is a floating average of the last x seconds, with x being
the damping factor. The display thus takes x seconds to fully respond to flow rate changes.
If
Swift damping
is deactivated, the damping is calculated as first order low-pass filter, i.e. changes of measured values
become effective in form of an exponential time course in the measuring result.
• Select the menu item
Swift damping
.
• Select
On
to activate the swift damping. Select
Off
to deactivate it.
• Select
Default
(
Swift damping
=
On
) if no customized inputs are to be made.
• Press ENTER.
Special functions\Measurement\Measurement settings
Special functions\Measurement\Measurement settings\Multi-point calibration
Special functions\Measurement\Measurement settings\Swift damping