13 User modes
13.2 ExpertUser mode
FLUXUS G831ST-HT
2021-06-01, UMFLUXUS_G831ST-HTV1-0EN
108
13.2.2
Threshold value for valid measurement
• Select the menu item
Options
in the program branch
Special settings
.
• Press ENTER until the menu item
Crest factor threshold
is displayed.
• Enter the threshold value (default: 6).
• Press ENTER.
13.2.3
Profile correction
It is possible to select the following versions for the calculation of the fluid mechanics calibration factor k
Re
:
•
kRe 1.0
: profile correction (previous version)
•
kRe 2.0
: improved profile correction (current version, default)
The following steps are necessary to set the profile correction:
Selection of the version
• Select the menu item
Special functions
in the program branch
Measurement settings
.
• Press ENTER until the menu item
Profile correction
is displayed.
• Select a list item (default:
kRe 2.0
).
• Press ENTER.
Measuring point specific profile correction
In special cases a measuring point specific profile correction can be used.
• Select the program branch
Options
.
• Press ENTER.
• Select a channel for which the profile correction is to be set.
• Select the menu item
Special settings
.
• Press ENTER until the menu item
Profile correction
is displayed.
• Select
Off
to deactivate the profile correction for the channel.
• Select
Default
to use the global setting from the program branch
Special functions
for the profile correction.
• Select
Customized
to use a measuring point specific profile correction.
• Press ENTER.
In case the list item
Customized
has been selected, the parameters of the measuring point specific profile correction will
now be displayed. The parameters of the profile correction are preferably transmitted to the transmitter via the service
interface but can also be entered here.
Notice!
If the limit of the flow velocity
+Flow velocity limit
is too low or
-Flow velocity limit
is too high, a
measurement might be impossible because most of the measured values will be marked as invalid.
Options\Special settings\Crest factor threshold
Notice!
A reduced threshold value allows for measurements under difficult conditions, but increases the risk of measurement
errors.
Special functions\Measurement\Measurement settings\Profile correction
Options\Special settings\Profile correction