16 Advanced functions
FLUXUS G809
16.1 Totalizers
133
UMFLUXUS_G809V2-2EN, 2020-06-25
16
Advanced functions
16.1
Totalizers
There are 2 totalizers, one for the positive and the other for the negative flow direction. The unit of measurement used for
totalizing corresponds to the volume or mass unit selected for the physical quantity.
The totalizer values can be displayed with up to 11 places, e.g.,
74890046.03
. For the adjustment of the decimal places
(max. 4), see section 17.7.
Display of the totalizer
• Press key
during the measurement to scroll through the upper line and display the totalizers.
The totalizer value will be displayed in the upper line (here: the volume which has passed the measuring point in positive
flow direction since the activation of the totalizers).
• Press ENTER while a totalizer is displayed to toggle between the display of a totalizer for both flow directions.
• Select the command
→Clear totalizer
in the upper line to reset the flow totalizers to zero.
• Press ENTER.
Automatic display toggling
The automatic toggling of the totalizer display between positive and negative flow direction can be set.
• Enter a time interval between 0 (off) and 5 s.
• Press ENTER.
Totalizer storing
It is possible to store the currently displayed totalizer only or to store one value for each flow direction.
• Select
Special Funct.\SYSTEM settings\Storing\Quantity Storage
.
• Press ENTER.
• If
one
is selected, only the value of the totalizer currently displayed will be stored. This can apply for the positive and
negative totalizer.
• If
both
is selected, the values of the totalizers for both flow directions will be stored.
• Press ENTER.
Totalizer behavior after the measurement is stopped
The totalizer behavior when the measurement is stopped or after a reset of the transmitter is set in the menu item
Special Funct.\SYSTEM settings\Measuring\Quantity recall
.
• If
on
is selected, the values of the flow totalizers will be stored and used for the next measurement.
• If
off
is selected, the flow totalizers will be reset to zero.
• Press ENTER.
Special Funct.\SYSTEM settings\Measuring\Toggle totalizer
Special Funct.\SYSTEM settings\Storing\Quantity Storage
Special Funct.\SYSTEM settings\Measuring\Quantity recall
A:Volume flow
54.5
m3/h
A:
32.5
m3
54.5
m3/h