18 Settings
FLUXUS G801
18.3 Setting of the standard conditions for gas measurements
157
UMFLUXUS_G801V1-0EN, 2020-06-26
18.3
Setting of the standard conditions for gas measurements
• Select the menu item
Gas-Measuring
.
• Press ENTER.
This display will only be indicated if the gas measurement has been activated in the menu item
Special Funct.\
SYSTEM settings\Measuring
.
• Enter the pressure for the local standard conditions.
• Press ENTER.
• Enter the temperature for the local standard conditions.
• Press ENTER.
18.4
Libraries
The internal database of the transmitter contains parameters for pipe and lining materials as well as for fluids.
The material and fluid scroll lists displayed in the program branch
Parameter
can be arranged. Shorter scroll lists make
work more effective.
• Select the menu item
Libraries
.
• Press ENTER.
• Select
Material list
to edit the material scroll list or
Medium list
to edit the fluid scroll list.
• Select
go back
to return to the menu item
SYSTEM settings
.
• Press ENTER.
• Select
factory
if all materials/fluids of the internal database are to be displayed in the scroll list. An already existing
customized scroll list will not be deleted but only deactivated.
• Select
user
to activate the customized scroll list.
• Press ENTER.
If
user
is selected, the material or fluid scroll list can be edited.
• Select
End of Edit
to stop editing.
• Press ENTER.
• Select
yes
to store all changes of the scroll list, or
no
to quit the menu item without storing.
• Press ENTER.
Special Funct.\SYSTEM settings\Gas-Measuring
Special Funct.\SYSTEM settings\Gas-Measuring\Norm/Std press.
Special Funct.\SYSTEM settings\Gas-Measuring\Norm/Std temper.
Special Funct.\SYSTEM settings\Libraries
Special Funct.\SYSTEM settings\Libraries\Material list
Special Funct.\...\Material list\user\Show list
Special Funct.\...\Material list\user\End of Edit
Special Funct.\...\Material list\user\Save List?