7 Connection
7.1 Transducers
FLUXUS F808
2020-06-25, UMFLUXUS_F808V2-2EN
70
7.1.11.2 Connection with potential separation
If earthing on the same potential cannot be ensured e.g., in measurement arrangements with long extension cables, the
extension cable and the junction box have to be electrically insulated from each other. The junction box and the transduc
-
ers have to be on the same potential. By doing this, no compensation currents can flow to the transmitter via the extension
cables.
For measurement arrangements where the junction box and the transducers have to be electrically insulated from each
other, see the document TIFLUXUS_GalvSep.
• Remove the blind plug for the connection of the extension cable.
• Open the cable gland of the extension cable. The compression part remains in the cap nut.
• Push the extension cable through the cap nut, the compression part and the basic part.
• Insert the extension cable into the junction box.
• Prepare the extension cable.
• Cut the outer shield and brush it back.
• Pull the extension cable back until the back brushed external shield is below the shield terminal, see Fig. 7.17. The ex
-
tension cable has to remain completely insulated up to the shield terminal.
• Screw the gasket ring side of the basic part into the junction box.
• Fix the cable gland by screwing the cap nut onto the basic part.
Fig. 7.16: Connection of the extension and transducer cable to the junction box
1 – extension cable
2 – external shield brushed back
3 – cap nut
4 – compression part
5 – basic part
6 – connection of the extension cable
7 – connection of the transducer cable
Important!
Observe the max. permissible voltage of 60 V DC between the earth potentials.
cable gland