7 Connection
FLUXUS F808
7.1 Transducers
55
UMFLUXUS_F808V2-2EN, 2020-06-25
7.1
Transducers
The transmitter has 1 cable gland for the connection of the transducers.
It is recommended to run the cables from the measuring point to the transmitter before connecting the transducers to
avoid load on the connectors.
Transmitter FLUXUS F808**-A1
7.1.1
Connection of the transducer cable to the transmitter
7.1.1.1
Transducer cable with plastic cable jacket and stripped cable ends
• Remove the blind plug for the connection of the transducer cable.
• Open the cable gland of the transducer cable. The compression part remains in the cap nut.
• Push the transducer cable through the cap nut and the compression part.
• Prepare the transducer cable.
• Shorten the external shield and brush it back over the compression part.
• Screw the sealing ring side of the basic part into the transmitter housing.
• Insert the transducer cable into the housing.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the transducer cable to the terminals of the transmitter, see Fig. 7.3 and Tab. 7.1.
Notice!
If transducers are replaced, the sensor module also has to be replaced, see section 7.1.12.
Fig. 7.2:
Connection of the transducers to the transmitter
1 – transducers
Important!
The degree of protection of the transmitter is only ensured if all cables are tightly fitted using cable glands and the
housing is firmly screwed.
Notice!
For good electromagnetic compatibility (EMC), it is important to ensure good electrical contact between the external
shield and the cap nut (and thus the housing).
ARS
AR
AV
AVS
L(+)
N(-)
PE
KL4
KL1
KL2
KL3
1
ARS
AR
AV
AVS