16 Advanced functions
16.7 Uncorrected flow velocity
FLUXUS F808
2020-06-25, UMFLUXUS_F808V2-2EN
132
16.7
Uncorrected flow velocity
For special applications, the uncorrected flow velocity might be of interest.
The profile correction for the flow velocity is activated in
Special Funct.\SYSTEM settings\Measuring\Flow
Velocity
.
• Select
normal
to display and output the flow velocity with profile correction.
• Select
uncorr.
to display and output the flow velocity without profile correction.
• Press ENTER.
If
uncorr.
is selected, each time the program branch
Measuring
is selected it will be requested whether the profile
correction is to be used or not.
If
no
is selected, the profile correction will be switched off. All physical quantities will be calculated with the uncorrected
flow velocity.
During the measurement, the designation of the physical quantity is displayed in capital letters to indicate that the value is
uncorrected.
• Press ENTER.
If
yes
is selected, the uncorrected flow velocity will only be used if the flow velocity is selected as physical quantity in the
program branch
Output Options
.
All other physical quantities (volumetric flow rate, mass flow rate, etc.) will be determined with the corrected flow velocity.
During the measurement, the designation of the physical quantity "flow velocity" is displayed in capital letters to indicate
that the value is uncorrected.
• Press ENTER.
In both cases, the corrected flow velocity can also be displayed.
Press key
to scroll until the flow velocity is displayed. The uncorrected flow velocity is marked with an "
U
".
Uncorrected flow velocities transmitted to a PC are marked with
uncorr.
16.8
FastFood mode
The FastFood mode allows to measure high dynamic flows. A continuous adaptation to changing measuring conditions is
only partially realized in the FastFood mode.
• The sound speed of the fluid is not updated. The last measured value of the sound speed before toggling to the
FastFood mode is used.
• The outputs can still be used. They are actualized every 100 ms independently from the storage rate.
• The measured values are stored with the storage rate of the FastFood mode, see 16.8.2.
• The FastFood mode has to be enabled and activated.
Special Funct.\SYSTEM settings\Measuring\Flow Velocity
PROFILE CORR.
>NO<
yes
FLOW VELOCITY
2.60
m/s
PROFILE CORR.
>no<
YES
Flow Velocity
*U
24
m/s