7 Connection
7.1 Transducers
FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
46
7
Connection
7.1
Transducers
It is recommended to run the cables from the measuring point to the transmitter before connecting the transducers to
avoid load on the connectors.
Warning!
Installation, connection and start-up by unauthorized and unqualified personnel
This may result in personal or material damage or dangerous situations.
→ Any work on the transmitter has to be carried out by authorized and qualified personnel.
Warning!
Touching live parts
Electric shock or arc faults can led to severe injuries. The measuring instrument can be damaged.
→ Prior to any work on the transmitter (e.g., installation, dismounting, connection, start-up), the transmitter
has to be disconnected from the power supply. It is not sufficient to remove the internal fuse of the
instrument (see section 7.2).
Caution!
Safety and accident prevention regulations for electrical systems and equipment
Failure to observe these regulations may lead to severe injuries.
→ Observe the safety and accident prevention regulations for electrical systems and equipment.
Notice!
If transducers are replaced, the transducer module also has to be replaced (see section 7.1.5).
Fig. 7.1:
Connection of the transducers to the transmitter
1 – transducer connection
AV
AV
S
AGN
ARS
AR
AV
AV
S
AGN
ARS
AR
1
Содержание FLUXUS F502BT
Страница 1: ...FLUXUS F502TE UMFLUXUS_F502TEV1 5EN Operating instruction...
Страница 6: ...Table of contents FLUXUS F502TE 2020 05 29 UMFLUXUSF502TEV1 5EN 6...
Страница 144: ...19 Thermal energy rate measurement 19 4 Steam in inlet FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 144...
Страница 160: ...Annex D Conformity declarations FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 160 D Conformity declarations...
Страница 161: ......
Страница 162: ......