3 General principles
FLUXUS F502TE
3.2 Measurement arrangements
13
UMFLUXUS_F502TEV1-5EN, 2020-05-29
3.2
Measurement arrangements
3.2.1
Terms
Sound path
The distance covered by the ultrasonic signal after crossing the pipe once. The number of the sound paths is:
• odd if the measurement is carried out in diagonal arrangement
• even if the measurement is carried out in reflection arrangement
Beam
The path covered by the ultrasonic signal between the transducers, i.e., the transducer emitting the ultrasonic signal and
the transducer receiving it. One beam consists of 1 or several sound paths.
Fig. 3.3:
Transit time difference Δt
1 – signal in the flow direction
2 – signal against the flow direction
Diagonal arrangement
Reflection arrangement
The transducers are mounted on opposite sides of the pipe.
The transducers are mounted on the same side of the pipe.
Fig. 3.4:
Diagonal arrangement with 1 beam and
3 sound paths
Fig. 3.5:
Reflection arrangement with 1 beam and
4 sound paths
1
2
Δt
Содержание FLUXUS F502BT
Страница 1: ...FLUXUS F502TE UMFLUXUS_F502TEV1 5EN Operating instruction...
Страница 6: ...Table of contents FLUXUS F502TE 2020 05 29 UMFLUXUSF502TEV1 5EN 6...
Страница 144: ...19 Thermal energy rate measurement 19 4 Steam in inlet FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 144...
Страница 160: ...Annex D Conformity declarations FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 160 D Conformity declarations...
Страница 161: ......
Страница 162: ......