16 Advanced functions
FLUXUS F502TE
16.7 Modification of the limit for the inner pipe diameter
115
UMFLUXUS_F502TEV1-5EN, 2020-05-29
• Activate or deactivate the auto-snap function.
• Press ENTER.
• Activate or deactivate the snap ringbuffer.
• Press ENTER.
Delete snaps
• Select
Clear Snaps
.
• Press ENTER.
Read snaps
• Select
Snaps ->Rs232
.
• Press ENTER.
Activation of the snap function
• In order to activate the snap function, press
until
DSPSNAP/VOLTAGE
is displayed in the upper line.
• Press ENTER.
16.7
Modification of the limit for the inner pipe diameter
It is possible to modify the lower limit of the inner pipe diameter for a given transducer type.
• Enter the HotCode
071001
(see section 8.4).
• Enter the lower limit of the inner pipe diameter of the displayed transducer type.
• Press ENTER to select the next transducer type.
Special Funct.\...\DSP-SignalSnap\AutoSnap
Special Funct.\...\DSP-SignalSnap\Snap ringbuffer
Special Funct.\SYSTEM settings\Signal snap\DSP-SignalSnap\Clear Snaps
Special Funct.\SYSTEM settings\Signal snap\DSP-SignalSnap\Snaps ->Rs232
Notice!
If a transducer is used below its recommended inner pipe diameter, a measurement might be impossible.
DNmin Q-Sensor
15
mm
Содержание FLUXUS F502BT
Страница 1: ...FLUXUS F502TE UMFLUXUS_F502TEV1 5EN Operating instruction...
Страница 6: ...Table of contents FLUXUS F502TE 2020 05 29 UMFLUXUSF502TEV1 5EN 6...
Страница 144: ...19 Thermal energy rate measurement 19 4 Steam in inlet FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 144...
Страница 160: ...Annex D Conformity declarations FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 160 D Conformity declarations...
Страница 161: ......
Страница 162: ......