14 Data logger
FLUXUS F502TE
14.4 Measurement with activated data logger
103
UMFLUXUS_F502TEV1-5EN, 2020-05-29
Totalizer storing
It is possible to store the currently displayed totalizer only or one value for each flow direction.
• Select
one
to store the currently displayed totalizer value only. This can apply for the positive and negative totalizer.
• Select
both
to store the totalizer values for both flow directions.
• Press ENTER.
Signal amplitude storing
• Select
on
to store the amplitude of the measured signal together with the measured values.
• Press ENTER.
Fluid sound speed storing
• Select
on
to store the fluid sound speed together with the measured values.
• Press ENTER.
Diagnostic values storing
• Select
on
to store the diagnostic values together with the measured values.
• Press ENTER.
14.4
Measurement with activated data logger
• Start the measurement.
• Enter the number of the measuring point.
• Press ENTER.
If the menu item
Output Options\Store Meas.Data
is activated and the menu item
Special Funct.\SYSTEM
settings\Ringbuffer
is deactivated, a message will be displayed as soon as the data logger is full.
• Press ENTER.
The error message will be displayed periodically.
Notice!
Storage mode
=
average
The average of the physical quantity and other quantities assigned to the measuring channel, will be calculated.
If the storage rate < 5 s is selected,
sample
will be used.
If no average could be calculated over the complete storage interval, the value will be marked as invalid. The ASCII
file of the stored data will contain
???
for invalid average values and
?UNDEF
instead of invalid temperatures.
Special Funct.\SYSTEM settings\Storing\Quantity Storage
Special Funct.\SYSTEM settings\Storing\Store Amplitude
Special Funct.\SYSTEM settings\Storing\Store c-Medium
Special Funct.\SYSTEM settings\Storing\Store diagnostic
Measuring\Channel\Meas.Point No.
DATA LOGGER IS FULL!
Содержание FLUXUS F502BT
Страница 1: ...FLUXUS F502TE UMFLUXUS_F502TEV1 5EN Operating instruction...
Страница 6: ...Table of contents FLUXUS F502TE 2020 05 29 UMFLUXUSF502TEV1 5EN 6...
Страница 144: ...19 Thermal energy rate measurement 19 4 Steam in inlet FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 144...
Страница 160: ...Annex D Conformity declarations FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 160 D Conformity declarations...
Страница 161: ......
Страница 162: ......