13 Outputs
FLUXUS F401
13.2 Activation of a binary output as pulse output
85
UMFLUXUS_F401V2-1EN, 2020-06-12
• Select
Opto-Relay ON
in the scroll list
Output Test
to test the energized state of
the output.
• Press ENTER. Measure the resistance at the output. The value has to be low ohmic.
• Select
yes
to repeat the test,
no
to return to the menu item
SYSTEM settings
.
• Press ENTER.
13.2
Activation of a binary output as pulse output
A pulse output is an integrating output which emits a pulse when the volume or the mass
of the fluid which has passed the measuring point reaches a given value (
Pulse Value
).
The integrated quantity is the selected physical quantity. Integration is restarted as soon
as a pulse is emitted.
• Press ENTER until
Pulse Output
is displayed. Select
yes
to activate the output.
• Press ENTER.
This error message will be displayed if the flow velocity is selected as the physical quantity.
The use of the pulse output is not possible in this case because the integration of the flow
velocity does not result in a reasonable value.
Special Funct.\SYSTEM settings\Proc. outputs\...\B1 Output Test\
Opto-Relay ON
Special Funct.\...\B1 Output Test\B1=ON\Again?
Notice!
The menu item
Pulse Output
will only be indicated in the program branch
Output Options
if a pulse output has been installed.
Output Options\...\Pulse Output
Output Options\...\Pulse Output\NO COUNTING