16 Advanced functions
FLUXUS F401
16.9 Diagnosis with the help of the snap function
111
UMFLUXUS_F401V2-1EN, 2020-06-12
Press key
to scroll until the flow velocity is displayed. The uncorrected flow velocity is
marked with an "
U
".
Uncorrected flow velocities transmitted to a PC are marked with
uncorr.
16.9
Diagnosis with the help of the snap function
By means of the snap function it is possible to store measuring parameters which are
useful for the evaluation of measuring results or diagnosis purposes.
• Select
Special Funct.\SYSTEM settings\Signal snap
.
• Press ENTER.
Snap memory settings
• Select
on
to activate the snap function and
off
to deactivate it.
• Press ENTER.
• Select
Install Snap
.
• Press ENTER.
• Enter the number of the snap memory storage space.
• Press ENTER.
• Activate or deactivate the auto-snap function.
• Press ENTER.
Special Funct.\SYSTEM settings\Signal snap
Special Funct.\SYSTEM settings\Signal snap\DSP-SignalSnap
Special Funct.\...\DSP-SignalSnap\Install Snap
Special Funct.\...\DSP-SignalSnap\Install Snap\Snap-Memory
Special Funct.\...\DSP-SignalSnap\AutoSnap
Flow Velocity
*U
54.5
m/s