Flashback Technologies CipherOx CRI M1 Скачать руководство пользователя страница 22

Page 

22

 of 

24

 

subject provided the basis for building models to estimate 

CRI, the proportion of remaining volume loss 

the individual can tolerate before  collapse (Equation 1)

. A robust  CRI estimate requires analysis of the 

continuous  pulsatile  waveform  signal  and  comparison  to  a  reference  database.  The  estimated  CRI  is 
determined by the most similar signal in the database. 

Clinical Validation of CRI 

To  characterize  the  relationship  of  CRI  to  significant  blood  loss  and  to  support  regulatory  clearance, 
Flashback Technologies conducted a prospective validation study that enrolled 42 healthy volunteers (19 
to 36) undergoing stepwise removal and replacement of approximately 20% of total blood volume while 
in the supine position at rest (non-motion). The methodology and preliminary results from the first 20 
subjects have been reported by (Convertino, Howard, et al. 2015) and the full dataset by (S. Moulton, et 
al. 2017). 
Results from the dataset of 42 volunteers demonstrate: 

 

Trending Intravascular Volume Changes: 32 of the 42 subjects completed the full protocol. For 
these subjects, the correlation between CRI and blood volume loss is greater than 0.9 (p<0.05). 

o

 

Correlation  between  average  CRI  values  (at  constant  intravascular  volume)  and 
intravascular volume loss is greater than 0.94 (95% confidence 0.92 to 0.96). 

o

 

Correlation  between  one-minute  average  CRI  and  intravascular  volume  loss  is  greater 
than 0.9 (95% confidence 0.87 to 0.93). 

 

In this population ‘normal’ CRI values (obtained from all 42 subjects) before intravascular volume 

loss were significantly above 0.7 (see box and whisker plots in Figure 13, and Figure 14). 

o

 

Before intravascular volume loss 95% of CRI values were above 0.75 and 98% of CRI values 
were above 0.7. 

o

 

Before  intravascular volume  loss  the mean  CRI  value  was 0.9  (95%  confidence  0.89  to 
0.91) with a standard deviation of 0.077 (95% confidence 0.053 to 0.078). 

 

At 20 % intravascular blood loss volume, CRI values (obtained from the 32 subjects that completed 
the full protocol) were significantly lower than those before blood loss (see box and whisker plots 
in Figure 13 and Figure 14). 

o

 

At 20% intravascular blood loss volume the mean CRI value was 0.6 (95% confidence 0.57 
to 0.63) with a standard deviation of 0.17 (95% confidence 0.16 to 0.17). 

 

When comparing one-minute average CRI values before blood loss to those at 20% intravascular 
blood loss volume (for the 32 subjects that completed the full protocol), the average drop was 
0.35 (95% confidence 0.29 to 0.4) with a standard deviation of 0.15 (95% confidence 0.11 to 0.2). 

 

CRI  values  when  subjects  were  experiencing  symptoms  associated  with  hemodynamic 
decompensation  were  significantly  lower  than  when  subjects  had  no  symptoms  (see  box  and 
whisker plots in Figure 13 and Figure 14). 

o

 

Seventeen percent of the subjects (7/42) experienced 

‘Hemodynamic

 

Decompensation’ 

during conduct of the study. 

 

As  evidenced  by  a  drop  in  systolic  BP  from  119  mmHg  to  <80  mmHg  and  a 
significant  drop  in  Mean  Arterial  Pressure  (average  decrease  in  MAP  was  43 
mmHg; average MAP prior to intravascular volume loss = 93 mmHg, range 91-95; 
average MAP at the point of hemodynamic decompensation was 50 mmHg, range 
44-58). 

o

 

During  symptoms  the  mean  CRI  value  was  0.15  (95%  confidence  0.12  to  0.17)  with  a 
standard deviation of 0.08 (95% confidence 0.05 to 0.09). 

o

 

During symptoms more than 97% of CRI values were below 0.3, while when no symptoms 
were present 98% of CRI values were above 0.3. 

 

Содержание CipherOx CRI M1

Страница 1: ...shback M1 Compensatory Reserve Monitor Handheld with Flashback s Patented CRI Technology Instructions for Use Version 1 0 2 May 2020 Flashback Technologies Inc 1 Little West 12th Street Suite 505 New...

Страница 2: ...o Software Symbols and Parameters 10 Parameters 10 Graphical Interface Symbols 10 Gauge 11 Trend 11 Sensor Status 11 Screen Status 12 Battery Status 12 Adjusting the Screen Brightness 12 Powering off...

Страница 3: ...x CRI 19 Estimating CRI from Pulsatile Waveforms 20 Performance of CRI 20 Summary of Clinical Study 21 Background Information Regarding Use of CRI 21 Compensatory Reserve 21 Development of the Compens...

Страница 4: ...ients with a finger thickness of 0 3 to 1 in hospital and pre hospital settings CRI trends with changes in intravascular volume relative to the individual patient s response to hypovolemia and should...

Страница 5: ...d replace the sensor Do not use if any component of the device is visibly damaged Do not use if the battery door is broken or fails to close tightly as this may allow ingress or ESD damage to the devi...

Страница 6: ...ying it to a patient The Model M1 is non sterile Do not sterilize autoclave or immerse the Model M1 in liquid Do not pour or spray any liquids into the device Do not use caustic or abrasive cleaning a...

Страница 7: ...patient isolation from electrical shock UL Mark for Canada and the United States with respect to electric shock fire and mechanical hazards only in accordance with IEC 60601 1 AAMI ANSI 60601 1 and C...

Страница 8: ...angle up to 60 from the vertical and ingress of solid foreign objects greater than or equal to 2 5 mm in diameter per IEC 60529 Do not expose the device to a magnetic resonance MR environment Conventi...

Страница 9: ...he left side of the M1 with the Nonin logo facing up see Figure 2 Figure 2 Attaching the Nonin Sensor Using the Nonin Articulated Clip Sensor 1 Insert a finger preferably the index middle or ring fing...

Страница 10: ...ually available within 5 to 20 seconds of sensor placement Figure 4 Power on the M1 Guide to Software Symbols and Parameters Parameters HR Heart Rate SpO2 Functional oxygen saturation of arterial hemo...

Страница 11: ...ee Figure 6 If the sensor is not connected or there is no signal the trend will continue to move leaving empty space in the graph in order to indicate time has passed since data has been collected Fig...

Страница 12: ...battery Less than 20 minutes of device use may remain The battery status indicator is only valid when an alkaline battery is used Adjusting the Screen Brightness The screen starts up in Normal brightn...

Страница 13: ...displayed as _ Cleaning the Reusable Sensor Cautions Clean the sensor before applying it to a new patient Unplug the sensor from the M1 before cleaning Do not sterilize autoclave or immerse the senso...

Страница 14: ...household bleach containing less than 10 sodium hypochlorite Do not sterilize autoclave or immerse the M1 in liquid of any kind Do not pour or spray any liquids onto the M1 Do not sterilize with EtO D...

Страница 15: ...Page 15 of 24 Nonin OEM III Specifications The M1 utilizes the internal Nonin OEM III module to generate SpO2 measurements...

Страница 16: ...5 volt AA battery CRI is not displayed Ensure snug connection between M1 and sensor OR Ensure Sensor Status state bottom left corner of M1 display shows a completely green outline and does not displa...

Страница 17: ...Group 1 The M1 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11...

Страница 18: ...50 DC 27 450 FM 1 kHz Sine 5 Hz Deviation 28 710 745 780 Pulse 217 Hz 50 DC 9 810 870 930 Pulse 18 Hz 50 DC 28 1720 1845 1970 Pulse 18 Hz 50 DC 28 2450 28 5240 5500 5785 9 M1 Compliance The M1 is not...

Страница 19: ...l the patient is already near collapse and at risk for poor tissue perfusion progressive acidosis and sudden unexpected hemodynamic decompensation The Compensatory Reserve Index CRI provides a single...

Страница 20: ...case CRI is based on the IR PPG signal from the Nonin OEM III 8000AA Performance of CRI CRI is a new physiological monitoring parameter which trends with intravascular volume and can be used by clinic...

Страница 21: ...tudied and observed how these parameters change in the setting of acute blood loss have long assumed that hypotension and other signs and symptoms of hemorrhagic shock mark the beginning of circulator...

Страница 22: ...Figure 14 o Before intravascular volume loss 95 of CRI values were above 0 75 and 98 of CRI values were above 0 7 o Before intravascular volume loss the mean CRI value was 0 9 95 confidence 0 89 to 0...

Страница 23: ...on CRI estimates 95 confidence Linear mixed model analysis with CRI as the outcome and intravascular volume changes as the predictor support the conclusion that CRI trends with changes in intravascul...

Страница 24: ...ms CRI estimates are based on reference data collected in ideal conditions Progressive declines to low CRI values indicate measurements based on subjects in clinical studies exhibiting symptoms of inc...

Отзывы: