18
R-Motion
®
File systems selection
File selection
Stand-by interface
Operation interface
4.5.2
FKG File Systems parameters availability table
Parameter
FKG File System
Set
Working
Speed
Set
Torque
Limit
Set
Apical
Action
Set Auto
Start
Set Auto
Stop
Set Flash
Bar
Position
Set
Apical
Slow
Down
Set
Forward
Angle
Set
Reverse
Angle
RACE EVO
All Files
YES
YES
YES
YES
YES
n/a
n/a
XP-endo Treatment
Shaper
YES
YES
YES
YES
YES
n/a
n/a
XP-endo Treatment
Finisher
YES
YES
YES
YES
YES
n/a
n/a
XP-endo Retreatment
DR1
YES
YES
YES
YES
YES
n/a
n/a
XP-endo Retreatment
Shaper
YES
YES
YES
YES
YES
n/a
n/a
XP-endo Retreatment
Finisher R
YES
YES
YES
YES
YES
n/a
n/a
R-Motion
All Files
YES
YES
YES
YES
NO
4.6
Set device parameters
To set the device parameters:
1.
To access the device parameters from the power OFF state, hold down the Setting button
“P” and press Main button.
2.
To select the desired parameter, press Setting button “P”
3.
To set the parameter, press the Adjusting button “+”/“-” and the Main button to confirm.
4.6.1
Device parameters list
Auto Power OFF
In stand-by state, the device automatically turns OFF after the set timer.
The timer can be adjusted from 3 minutes to 30 minutes (1 minute increments)
Auto Standby Scr
The display switches automatically back to the stand-by interface after the set timer.
The timer can be adjusted from 3 seconds to 30 seconds (1 second increments)
Dominant Hand
The device can be adjusted for left-handed or right-handed user (180° rotation of the
display).