Measuring
7.3.7
Measurements With Outlier Rejection Enabled
When making measurements with outlier rejection enabled:
the outlier measurements recognized by the instrument will appear on
the display and will be indicated acoustically (
two long audible signals will sound to indicated the outlier measure-
ment in addition to the audible signal for the measurement acquisition,
recognized outliers will not be integrated in the evaluation of the cur-
rent block or the current Application.,
only the signals for indicating the outlier measurements will sound if the
measurement acquisition signal has been disabled,
outlier measurements recognized in the auto-averaging mode will not
be integrated in the measurement calculation, and
only the audible signals indicating an outlier measurement will sound
when an outlier is recognized (a violation f the tolerance limits will not
be indicated).
Detail of the display
Explanation
and
will appear in front of the reading on
the display after a measurement has been taken
that has been recognized by the instrument as an
outlier,
Detail of the display
Explanation
Outlier!
appears briefly in the information lines of the dis-
play in addition to the audible signal if a past reading
with the number n (in the example, n = 1) is recognized
as an outlier measurement (thereafter, the measurement
just taken will appear).
Содержание FERITSCOPE FMP30
Страница 1: ...FERITSCOPE FMP30 Operator s Manual Coating Thickness Material Analysis Microhardness Material Testing ...
Страница 63: ...Page 58 Operator s Manual FERITSCOPE FMP30 Applications Fig 5 3 Example for linking of applications ...
Страница 235: ...Page 230 Operator s Manual FERITSCOPE FMP30 Glossary ...
Страница 240: ...Operator s Manual FERITSCOPE FMP30 Page 235 X Xn 79 ...
Страница 241: ...Coating Thickness Material Analysis Microhardness Material Testing ...