Page 52
Operator‘s Manual FERITSCOPE
®
FMP30
Applications
5.6.5
Unit
-ferrite content displayed as
-Martensit content. The measured
-ferrit
content in Fe% will be computed into the corresponding
-Martensit con-
tent. Further information see Chapter
-Martensit con-
tent.
5.6.6
Offset
The determined measured value is corrected by the defined offset value.
The offset value can be between -9999 and +9999 μm. The offset value
will always subtracted from the measured value. A positive Offset value de-
creases the displayed value, a negative Offset value increases the dis-
played value.
Key sequ./
Action
Detail of display
Explanation
MENU
Use
MENU
to start Settings.
Press the arrow key
, until “
Unit
” is high-
lighted on the display.
ENTER
Use
ENTER
to confirm the selection “
Unit
”.
or
Use
ENTER
to confirm the unit „
Fe%
“
or
Press the arrow key
to specify the unit
-
Martensit content „
M%
“.
ENTER
Press
ENTER
to confirm the selection/entry
and return to the menu to enter additional
settings.
Press
DEL
to return to the measurement
screen.
The instrument is ready to make measure-
ments.
Содержание FERITSCOPE FMP30
Страница 1: ...FERITSCOPE FMP30 Operator s Manual Coating Thickness Material Analysis Microhardness Material Testing ...
Страница 63: ...Page 58 Operator s Manual FERITSCOPE FMP30 Applications Fig 5 3 Example for linking of applications ...
Страница 235: ...Page 230 Operator s Manual FERITSCOPE FMP30 Glossary ...
Страница 240: ...Operator s Manual FERITSCOPE FMP30 Page 235 X Xn 79 ...
Страница 241: ...Coating Thickness Material Analysis Microhardness Material Testing ...