Page 28
Operator‘s Manual FERITSCOPE
®
FMP30
System Setup, Maintenance and Cleaning
3.3
Instrument: On/Off
3.3.1
Measurement Method of the Connected Probe
After powering up the instrument, the measurement method
[Ferrite]
ap-
pears on the display. The unit of measurement for displaying the readings
of the current Application is displayed next to it (
ning on Page 179 and 5.6.5 ‘Unit’, beginning on Page 52).
To avoid erroneous readings, no metallic objects must be in close
proximity to the probe tip when powering up the instrument.
The minimum distance is 25 mm.
Unit
Explanation
Fe % or
M%
-
ferrite content in ferrite percent or
-Martensit content
FN
-
ferrite content in ferrite numbers
If [Ferrite] flashes on the display, an Application has not yet been set up
using the connected probe. It is not possible to make measurements when
the display is flashing.
To make measurements, an Application must be set up using the con-
nected probe
(
Содержание FERITSCOPE FMP30
Страница 1: ...FERITSCOPE FMP30 Operator s Manual Coating Thickness Material Analysis Microhardness Material Testing ...
Страница 63: ...Page 58 Operator s Manual FERITSCOPE FMP30 Applications Fig 5 3 Example for linking of applications ...
Страница 235: ...Page 230 Operator s Manual FERITSCOPE FMP30 Glossary ...
Страница 240: ...Operator s Manual FERITSCOPE FMP30 Page 235 X Xn 79 ...
Страница 241: ...Coating Thickness Material Analysis Microhardness Material Testing ...