background image

5.5 cu.ft. Undercounter Refrigerator

 Getting To Know Your Product  |   11

Product Exterior – Rear

Access Port 

The Access Port (located on the back of the unit) is shipped 
sealed with a foam plug that must be removed prior to use. 
Additionally, the gray plastic knock-out must be removed. 
After inserting your device, we recommend reinserting the 
foam plug as snug and flush as possible to insulate against 
cooling loss. Putty may also be used to help reseal the area.

Evaporation Tray

The evaporation tray is located on the bottom rear of the unit. 

Product Interior

Interior Components

Air temperature sensor

External pass-through port

Adjustable shelving

Adjustable Shelving 

Your Isotemp undercounter refrigerator is shipped with wire 
ties holding the shelves in place. Remove the wire ties before 
using the refrigerator. You may place the shelving wherever 
you choose by removing and relocating the support clips.

Содержание Isotemp Series

Страница 1: ...5 5 cu ft Undercounter Refrigerator Isotemp Series Installation and Operation 329153H01 Revision A April 2018...

Страница 2: ...ormed by qualified service personnel Material in this manual is for informational purposes only The contents and the product it describes are subject to change without notice Fisher Scientific makes n...

Страница 3: ...otemp Refrigerator 5 Inspecting for Damage 5 Unboxing and Locating 5 Identifying all Parts 5 Installation 6 Leveling 6 Intended Use 6 Providing Ventilation 6 Power On and Power Cord Guidelines 6 Unint...

Страница 4: ...roduct Interior 11 Navigating Isotemp 5 5 Cu Ft Refrigerator Controls 12 System Alerts 12 System Alert Values 12 Reading the LCD Temperature Display 12 Access the Main Menu 12 Maintenance 16 Frequentl...

Страница 5: ...to clean interior aluminum walls Internal Chamber Capacity 5 5 cu ft 156 L Chamber Temperature Uniformity 3 C 5 4 F from center air temperature Cooling System Refrigeration system with single 125W th...

Страница 6: ...certified by an EPA recognized certification body to use less energy and cause fewer emissions that contribute to climate change ENERGY STAR and the ENERGY STAR mark are registered trademarks owned b...

Страница 7: ...function or injury Removal of the top cover for maintenance and other service should be performed only by factory authorized personnel CAUTION For Class B Unintentional Radiators This device complies...

Страница 8: ...Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including in...

Страница 9: ...our new product safely and easily Unboxing and Locating We recommend you move the box as close as possible to its final location before you begin to unpack This unit is not provided with handles there...

Страница 10: ...er refrigerator is equipped with a ventilation area just above the door handle Do not block or cover the venting area as this can impact the performance of the refrigerator Power On and Power Cord Gui...

Страница 11: ...wer source fails If you use a UPS backup take care to follow the directions and precautions accompanying that device Be careful not to plug into a wall outlet controlled by a wall switch to avoid acci...

Страница 12: ...run and the refrigerator will work to reach the default set point temperature Once set point is reached the fan speed will reduce to conserve energy The unit will reach air temperature setting rapidly...

Страница 13: ...s Avoid storing beverages or any non sample items like food in any unit used for sensitive inventory Not only is this a health hazard it also leads to more frequent opening of the door Note Isotemp un...

Страница 14: ...battery serves to support alerting the user in the event of a power loss Ventilation The Isotemp refrigeration system is a solid state system utilizing fans with moving parts The functioning of the f...

Страница 15: ...g and flush as possible to insulate against cooling loss Putty may also be used to help reseal the area Evaporation Tray The evaporation tray is located on the bottom rear of the unit Product Interior...

Страница 16: ...ator Controls 5 5 cu ft Undercounter Refrigerator Navigating Isotemp 5 5 Cu Ft Refrigerator Controls System Alerts System Alert Values Reading the LCD Temperature Display Access the Main Menu Color Di...

Страница 17: ...een that lists each alarm triggered Alarm Log The Alarm Log screen displays a history of all alarms Alarm Settings The Alarm Settings menu allows you to Select the high and low temperature alarms It a...

Страница 18: ...s the ability to coordinate the results of a third party external NIST calibrated reference thermometer with the temp displayed on the refrigerator Example Refrigerator air temperature displays as 3 5...

Страница 19: ...Cu Ft Refrigerator Controls 15 Model Information This displays the model number UI FW version UI FW date and TEC FW version of your unit System Information This displays the current settings which inc...

Страница 20: ...attachment Note Do not remove the top cover Do not attempt to clean the power supply located below the fan assembly You could damage your unit Cleaning the Condensate Tray The condensate tray is foun...

Страница 21: ...ontinuous power to the audible and visual alerts of power loss How long before I can use the product In a normal ambient environment your refrigerator should pull down to the target set point within a...

Страница 22: ...oor alarm keeps sounding Check to make sure the door is properly aligned in the frame by looking for uneven surfaces For example is the gap from the hinge to edge equal in width all the way across the...

Страница 23: ...shall apply only if Fisher Scientific verifies the applicable defect upon examination 5 Any defects or other damages resulting from shipment of the unit are not covered by this warranty Fisher Scienti...

Страница 24: ...verr smaltito o riciclato tramite queste medesime Ulteriori informazioni sulla conformit di Fisher Scientific con queste Direttive l elenco delle ditte di riciclaggio nel Vostro paese e informazioni...

Страница 25: ...sher com Italie 39 02 950 59 478 it fishersci com 39 800 823 162 unity it thermofisher com Finlande 358 0 9 8027 6280 fishersci fi 46 8 556 468 20 unity fi thermofisher com France 33 0 3 88 67 14 14 f...

Страница 26: ...Road Asheville NC 28804 United States Find out more at fishersci com 2018 Fisher Scientific Inc All rights reserved All trademarks are the property of Fisher Scientific and its subsidiaries unless oth...

Отзывы: