Chapter 4 | Calibration 

Chapter 4 
Calibration 

When the conductivity calibration is the only calibration 
performed, it will be applied to all measurement modes. 
When a TDS calibration is performed, it will only be applied 
to TDS measurements. When a salinity calibration is 
performed, it will only be applied to salinity measurements.  

The non-volatile meter memory retains the most recent, 
active calibration data when powered off. 

Perform a calibration with one to five conductivity 
standards, one to five TDS standards and one salinity 
standard. A maximum of one calibration point per range 
can be performed for conductivity and TDS calibrations. 

 

Conductivity 

Range 

Automatic Calibration Value 

Reference Temperature: 

15°C 

20°C 

25°C 

30°C 

r 1  0.00 - 20.00 μS 

none 

none 

none 

none 

r 2  20.01 - 300.0 μS  68.00μS  76.00μS  84.00μS  92.00μS 
r 3  300.1 - 9999 μS  1147μS  1278μS  1413μS  1548μS 
r 4 1.000 - 99.99 mS 10.48mS 11.67mS 12.88mS 14.12mS 
r 5 100.0 - 200.0 mS  92.5mS  102.1mS 111.8mS 121.8mS 

 

 

TDS 

Range 

r 1  0.00 - 10.00 ppm 
r 2  10.01 - 200.0 ppm 
r 3  200.1 - 9999 ppm 
r 4  1.00 - 99.99 ppt 
r 5  100.0 - 500.0 ppt 

 

When the conductivity electrode is replaced, it is best to 
clear the calibration log. Refer to the 

Clear Data

 section for 

detailed instructions.  

4.1 Conductivity 

Calibration 

4.1.1 Conductivity Calibration 
Overview 

For best results, periodic calibration with known, accurate 
and fresh conductivity standards is recommended. Always 
use fresh conductivity standards. Select a conductivity 

standard or standards that are closest to the expected 
sample conductivity value or values. Thoroughly rinse the 
conductivity electrode before calibration and between 
samples with deionized water.  

During the conductivity calibration, the meter will 
automatically recognize conductivity standard values of 84 
μS, 1413 μS, 12.88 mS and 111.8 mS (values at 25°C). 
The 

Temperature Compensation Coefficient

 setting in the 

Setup Menu will determine how the conductivity calibration 
standards are displayed during the calibration. 

 

If temperature compensation is set to Off: Once the 
calibration reading is stable, the meter will display 
the conductivity standard value at its measured 
temperature (conductance).  

 

If temperature compensation is set to On: Once the 
calibration reading is stable, the meter will display 
the temperature compensated conductivity standard 
value, based on the 

Reference Temperature

 

selected in the Setup Menu (15, 20, 25 or 30°C). 

If the conductivity reading does not correspond with an 
automatically recognized conductivity standard value 
(within 

±

40% window), once the conductivity reading is 

stable, the meter will display the theoretical value of the 
conductivity standard according to the 

Nominal Cell 

Constant

 value entered in the Setup Menu. The operator 

must manually enter the conductivity standard value at the 
measured temperature (conductance). 

During calibration, each conductivity standard value can be 
accepted or edited as needed. If the conductivity standard 
value needs to be manually adjusted, use the 

 or 

 key 

to edit the value of the conductivity standard at its 
measured temperature. 

After performing a one point conductivity calibration, the 
meter will apply the resulting calibration factor across all 
conductivity ranges. After performing a two, three, four or 
five point conductivity calibration, the meter will apply the 
resulting calibration factor of each conductivity standard to 
its applicable conductivity range and apply the most 
relevant calibration factor to any conductivity ranges for 
which a standard was not used during calibration. 

During the conductivity calibration, press the “Cal” key to 
save and end the calibration after accepting the last 
desired calibration point. For example, to perform a two 
point calibration, press the “Cal” key after accepting the 
second calibration point. This allows the operator to 
perform a conductivity calibration using one to five points 
for each calibration without having to program the number 
of calibration points prior to the calibration. 

Содержание Fisherbrand accumet AB330

Страница 1: ...Operations Manual Electrochemistry Meters 68X002128 Revision 01 September 2021 accumet AB330 Conductivity Meter User Manual...

Страница 2: ......

Страница 3: ...3 1 Setup Menu Scrolling 8 2 3 2 Setup Menu Value Changes 8 2 3 3 Mode Shortcut Key 8 2 3 4 Date and Time View Shortcut Key 8 2 4 Display Overview 9 2 5 Meter Connections 10 2 6 Electrode Information...

Страница 4: ...ue 19 3 2 16 Set Time Value 20 3 2 17 Audio Mode 20 3 2 18 Sleep Mode 20 3 2 19 Clear Data 21 3 2 20 Factory Reset 21 Chapter 4 22 Calibration 22 4 1 Conductivity Calibration 22 4 1 1 Conductivity Cal...

Страница 5: ...30 Meter Info 30 7 1 Meter Specifications 30 7 2 Meter Dimensions 31 7 3 Warranty 31 7 3 1 Return of Items 31 Chapter 8 32 Regulatory Compliance 32 8 1 European Union 32 8 2 Product Safety 32 8 3 Elec...

Страница 6: ...Note This symbol in the context of a Note indicates to follow the manufacturer specified instructions notes and requirements set out in the instruction manuals CAUTION Before using or maintaining thi...

Страница 7: ...ith US Japan Euro UK Singapore Australia New Zealand and China wall plug plates is included with the meter This universal power adapter is specifically for use with this meter Use of other power adapt...

Страница 8: ...ate and time setting Press to save a setting in the Setup Menu Press to accept a calibration point in the Calibration mode Press to power on the meter When the meter is on press to turn the backlight...

Страница 9: ...ter is in the Measure mode Memory Icon Indicates meter is in the Setup Menu and viewing the data log or calibration log Log Saved Icon Indicates meter is actively saving data to the data log or calibr...

Страница 10: ...purchase electrode separately or kitted with a conductivity electrode Refer to the electrode instructions for specific care and maintenance procedures The following are general recommendations for el...

Страница 11: ...saved data logs and calibration logs to the selected export device 4 Temperature Calibration or Manual Temperature Input When an ATC probe is connected use the Temp Calibration menu to perform a temp...

Страница 12: ...pressed the View Logs setup menu item is always shown first Press the or key to scroll through the data and calibration log options Press key Press key Press key Press key Press key Press the Enter ke...

Страница 13: ...he Enter key to view the display log For the calibration log press the or key to change the view from the date time screen to the calibration data screen Press or key For the calibration log press the...

Страница 14: ...d export device 3 2 4 Temperature Calibration or Manual Temperature Input 3 2 4a Temperature Calibration When an ATC probe is connected use the Temperature Calibration menu to perform a one point temp...

Страница 15: ...etup Menu list until the Measure Mode item is shown Press the or key to set the measure mode Press the Enter key to save the change 3 2 6 Nominal Cell Constant Set the nominal value for the cell const...

Страница 16: ...inear temperature compensation value enter a value of 0 0 to 10 0 C Press the or key to scroll through the Setup Menu list until the Temperature Compensation Coefficient item is shown Press the or key...

Страница 17: ...Export key to save a measurement to the data log and export Auto Read In the Auto Read measure mode press the Read key to start a measurement When the measurement is stable the checkmark stable icon i...

Страница 18: ...time interval Press the or key to scroll through the Setup Menu list until the Calibration Due Alarm item is shown Press the or key to adjust the alarm value a Press the key to increase the alarm valu...

Страница 19: ...he month value will flash Press the or key to adjust the month a Press the key to increase by one month Press and hold the key to quickly increase b Press the key to decrease by one month Press and ho...

Страница 20: ...the key to decrease by one minute Press and hold the key to quickly decrease Press the Enter key to save the change 3 2 17 Audio Mode Set the audible beep on or off The audible beep is used each time...

Страница 21: ...onfirm when the selected log is fully erased The meter will return to the measurement screen 3 2 20 Factory Reset Erase all data logs calibration logs and settings and return the meter to its factory...

Страница 22: ...ll determine how the conductivity calibration standards are displayed during the calibration If temperature compensation is set to Off Once the calibration reading is stable the meter will display the...

Страница 23: ...alue is shown press the Enter key to accept the conductivity standard value and proceed to the next calibration point To calibrate with another conductivity standard repeat steps 3 5 or to save and en...

Страница 24: ...ept the TDS standard value To calibrate with another TDS standard repeat steps 3 6 or to save and end the calibration press the Cal key The meter will display the calculated calibration factor value a...

Страница 25: ...T traceable thermometer Press the Menu key to enter the Setup Menu Press the or key to scroll through the Setup Menu list until the Temp Calibration item is shown The temperature value will blink indi...

Страница 26: ...eter is in the measurement mode 5 2 Conductivity Salinity and TDS Measurements Rinse the electrode with distilled water or appropriate solution and blot gently with a lint free tissue to remove excess...

Страница 27: ...saved with its associated date and time stamp To view the time stamp with seconds included the data log must be exported to a computer or printer The conductivity calibration log and TDS calibration l...

Страница 28: ...mode MODE COND or SALT or TDS Remote Command GETLOG CR Prints all logged measurement data If no data is logged returns to receive next command Output format is based on the Log Export Type setting Rem...

Страница 29: ...sing plates and temperature sensor submerged below the top of the sample solution o Verify cables do not have any damage o Perform a factor reset on the meter 6 2 Measurement Troubleshooting Reading i...

Страница 30: ...p to 9 F Features Display 5 backlit LCD Time and Date Yes Read Types Continuous Auto Read Timed Timed Interval 5 seconds to 60 minutes Hold Function Yes Data Log 500 data sets with time and date Calib...

Страница 31: ...epaired on a charged basis The warranty on your meter shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of...

Страница 32: ...ironment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communicat...

Страница 33: ...der entsorgt werden Mehr Informationen ber die Einhaltung dieser Anweisungen durch Thermo Scientific dieVerwerter und Hinweise die Ihnen n tzlich sein k nnen die Thermo Fisher Scientific Produkte zu i...

Страница 34: ...____________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________...

Страница 35: ..._______________ __________________________________________________ __________________________________________________ __________________________________________________ _______________________________...

Страница 36: ...40V 13 637 690 accumet RS232 cable for thermal compact printer 13 637 681 accumet paper for compact printer 2 pack 13 637 691 Find out more at fishersci com Distributed by Fisher Scientific Contact us...

Отзывы: