Quick Start 

The recorder has been preset to operate within the parameters of the of 
the Fisher Isotemp Refrigerator or Freezer in which it is installed.  A pen 
and chart have already been installed for your convenience.  All you 
need to do to start using your recorder is follow these quick start 
instructions: 
1. Remove the protective pen cap, and insert the cap on the pen cap 

holder. 

2. Press the ON/OFF key to turn the instrument on (Figure 2). 
3. The instrument will enter a 5 second warm-up phase, ending with the 

pen moving to the appropriate temperature reading. Under warm-up 
mode the green LED flashes until ready, when ready, LED will stay 
solid green and F or C will be shown on the display. 

4. Set the appropriate time by pressing the CHART ROTATION key. 

(Figure 2).  The chart will turn clockwise while the button is 
depressed.  Continue the rotation until the correct hour and day is 
referenced to the timing arrow (Figure 1). 
Note: Do not try to rotate the chart hub manually - doing so may 
damage the instrument. 

7. For battery back up insert a 9 volt battery.  The battery door is center 

bottom of dial plate. 

Specifications 

 

Temperature Accuracy:

 

±

0.5 % of full scale. 

 

Temperature Sensor:

 Bead wire probe, type K-thermocouple 

 

Recording Times:

 24 Hour, 7 Day 

 

Chart:

 6

 diameter (15.2 cm) 

 

Ambient Operating Conditions:

 32 to 122

°

F (0 to 50

°

C) 

0 to 90% RH (non-condensing) 

 

Alarms (

Not used on this model

):

 Maximum and minimum alarm 

contacts audio and visual alarms 

Note: 

Response time is slower when using battery power source

 

Keypad Operations 

On/Off 

The On/Off key turns the unit on and off. Alarm and calibration settings 
are restored to the most recent values when the unit is turned on.  If the 
unit is using AC power and power fails, the unit will switch to battery 
power without interrupting the recording capabilities.  If power fails when 
the battery is not present, the pen will remain on the chart in recording 
position. 

Pen Home (also Down Arrow) 

If the pen is located outside the chart, press the 
HOME key to move pen to recording position.  Press 
the HOME key with the pen in recording position, 
and the pen will move to outside of the chart and up 
the pen lift ramp (home).  The PEN HOME key is 
also used as a down arrow key to decrease values 
when setting alarms or calibration values. 

Chart Rotation (also Up Arrow) 

The CHART ROTATION key sets the chart to the 
desired time setting.  Press the CHART ROTATION 
key to rotate chart.  Continue pressing until the chart 
reaches the desired time setting.  The chart rotation 
will reach full speed after a couple seconds of 
pressing the key.  

Do NOT attempt to rotate the 

chart manually.  

The CHART ROTATION key is 

also used as an up arrow key to increase values 
when setting alarms or calibration values. 

Alarm (Not used on this model) 

The ALARM key turns the alarm on or off. When the 
alarm is on the display will alternate between a 
flashing A and F (in Fahrenheit mode) or A and C (in 
Celsius mode).  Holding the ALARM key for 2 
seconds will cause the unit to enter into the alarm 

set mode. For a visual alarm the display will read MIN if a minimum 
alarm is activated or MAX if the maximum alarm is activated.  The 
alarm condition will also trigger an audio alarm if the audio alarm is 
enabled (dip switch #8 is on).  The alarm will continue until the alarm is 
disabled, or until the alarm condition no longer applies. 

The PEN HOME and CHART ROTATION keys are used for setting 
alarms.  The PEN HOME key decreases the alarm value and the 
CHART ROTATION key increases the alarm value. 
1. To set the alarms press the ALARM key and hold for 2 seconds.  The 

temperature alarm set point will flash. 

2. Use the HOME key to decrease the minimum alarm setting and 

TIME key to increase the minimum alarm setting. 

3. Press ALARM key a second time and the temperature alarm set 

point will flash indication that you can now set the maximum alarm.  
Use the same procedure as in step #2 for setting the maximum alarm 
setting. 

4. When you have set the alarms press the ALARM key a third time to 

exit the alarm setting mode. 

5. To enable the ALARM, press the ALARM key. 