WinFTM
®
177
Control the XY(Z) Measuring Stage
Features of WinFTM BASIC and LIGHT
Travel to absolute or rela-
tive coordinates
You can enter absolute or relative coordinates to travel the
stage. The zero point of an operation can be specified indi-
vidually.
Laser light pointer
The red laser point points to the subsequent measurement
position in the retracted condition and thus significantly
simplifies a quick sample positioning
Point & Shoot
You can travel to the desired position by clicking the left
mouse button.
Point & Measure
You can travel to the desired position and perform a
measurement at this position by clicking the middle mouse
button.
Point & Save
You can travel to the desired position and save the related
coordinates by clicking the right mouse button.
Joystick at measuring
head
You can easily control the XY(Z) stage with the joystick at
the measuring head.
In addition, you can control the XY(Z) stage with the mouse.
Measurement point over-
view
The measurement point overview windows offer you a full
overview of all programmed points of the current measure-
ment coordinates and of the entire travel range.
Programming individual
coordinates
You can program arbitrary individual coordinates, on which
measurements will be performed.
Programming lines
You can define a start point and an end point of a line and
a number interim points. On these points measurements
will be performed.
Programming arrays
You can define two corner coordinates of an array and the
number of points within the array (columns and rows).
Furthermore you can define the sequence, in which the
points will be travelled. On these points, measurements will
be performed.
Programming patterns
You can define a freely programmable geometric pattern of
measurement points. You can replicate this pattern as often
as you want. Thus, you can travel the same points on
different specimens or on different parts of the same spec-
imen.
Programming offset
In addition to the measuring points (line, array or pattern)
you can define one or two reference points. Thus, you can
change specimens and make sure the measurements are
performed at the same coordinates, while the specimen is
not located at the same position.
Function
Description
Содержание FISCHERSCOPE X-RAY 5000 Series
Страница 18: ...18 FISCHERSCOPE X RAY Switching the Instrument on and off Communication between Instrument and WinFTM...
Страница 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Страница 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Страница 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Страница 112: ...Measurement device monitoring for the Fischerscope X RAY Long term monitoring 112 FISCHERSCOPE X RAY...
Страница 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Страница 124: ...124 FISCHERSCOPE X RAY Calibration...
Страница 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Страница 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Страница 182: ...182 WinFTM WinFTM SUPER For the Experienced X RAY User Comparative Overview with without WinFTM PDM...
Страница 183: ...WinFTM 183...