Functional Principle of the Instrument
Instrument Overview
2.3
Functional Principle of the Instrument
The following figure shows the principle structure of the instrument:
Fig. 2-2: Functional principle of the instrument
Functional principle
1. The x-ray tube generates the primary x-radiation (primary radiation). The electrically
heated cathode emits electrons. Accelerated by the applied high voltage to very high
speeds, the electrons bombard the anode material. This generates the primary
x-radiation.
2. The shutter serves as a safety device and closes the access of the primary x-radiation
to the measurement chamber, if needed.
X-ray tube
Cathode
Anode
Primary x-radiation
Aperture
Coating layer
Base material
X-ray fluorescence radiation
Spectrum
WinFTM
main window
Detector
Shutter
Содержание FISCHERSCOPE X-RAY 5000 Series
Страница 18: ...18 FISCHERSCOPE X RAY Switching the Instrument on and off Communication between Instrument and WinFTM...
Страница 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Страница 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Страница 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Страница 112: ...Measurement device monitoring for the Fischerscope X RAY Long term monitoring 112 FISCHERSCOPE X RAY...
Страница 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Страница 124: ...124 FISCHERSCOPE X RAY Calibration...
Страница 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Страница 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Страница 182: ...182 WinFTM WinFTM SUPER For the Experienced X RAY User Comparative Overview with without WinFTM PDM...
Страница 183: ...WinFTM 183...