Once the value has been selected, save it with
û
and
EIN
will appear again in the display.
Then leave the setting mode. Press
ÿ
until
Esc
and
the
û
appear. The current measured pressure is
shown again. The correct pressure unit is now illu-
minated to the right of this.
The display can only show up to ±1999.
Therefore in some cases it may not be
possible to select all stated pressure
units.
5.3.3
Display settings
Due to the fact that the DE44 has two independent
pressure measuring systems but only one display,
you can use the
dSP
parameter to select whether
you want to have only the first channel (
dSP
= 0.0),
only the second channel (
dSP
= 0.1) or both chan-
nels alternately (
dSP
= 0.2 to 99.9) displayed. From
0.2 s upwards, the value simply determines the re-
tention period (0.2 s to 99.9 s per channel). So that
it is clear which value is currently being displayed,
the symbols P1 for the 1st channel and P2 for the
2nd channel are shown accordingly on the left edge
of the display.
It is advisable for the other settings to in-
itially select a slow switching time (e.g. 3
s). This saves you the effort of having to
switch between P1 and P2 manually.
5.3.4
Zero point control and adjustment
Ensure that the device is not under any pressure (if
necessary, disconnect any pressure lines).
If the device does precisely indicate zero at this
point of time, parameter
OF1
enables you to adjust
the measuring value exactly to zero. To do so, you
have to set the measuring value indicated below
OF1
to zero.
Due to the fact that the DE44 has two independent
pressure measuring systems, there is also an offset
correction parameter
OF2
for the second measuring
channel. The function and effect is identical to
OF1
.
After zero-point adjustment, the pressure sensing
lines can be reconnected.
5.3.5
Damping and zero-point stabilising
If there are unsteady pressure readings at this point
of time or during operation, you can use parameters
DAM
and
NP
to stabilise the reading (and the output
signal).
The parameter
DAM
functions like a capillary throttle.
However, it only acts on the display, output signal
and switch points but not on the measuring cell it-
self. This parameter can be used to set the reaction
time to the pressure jumps. The value range is 0.0 s
to 100.0 s.
But with maximum attenuation, it will
take more than 2 minutes for the read-
ing to also reach zero after a pressure
jump from nominal pressure (100 %) to
zero!
In many cases, unsteady readings are not a prob-
lem during normal operating mode, but this is not
true for the idle state, i.e. if zero (differential) pres-
sure is expected.
In such situations, parameter
NP
can be applied. Its
value defines a measuring value range around ze-
ro. Within this range, the measuring value is set to
zero.
Example:
A value of 0.08 mbar
3
is entered for
NP
. In this
case all pressures within the range of -0.08
mbar to +0.08 mbar are set to zero. The read-
ing will only not indicate zero anymore if the
pressure exceeds these limits. However, the
pressure value and display do not correspond
to one hundred percent. The measuring pres-
sure and reading match again when the double
value, in this case 0.16 mbar, is reached again.
5.3.6
Setting the output signal
The transmitter output signal primarily depends on
the sensed pressure. However, you have the option
of adjusting the output signal to a large extent to
suit your requirements.
However, the basic measuring range
(indicated on the type label) and the
type of output signal (voltage / current)
are not variable.
Parameters
MA
(start of measuring range) and
ME
(end of measuring range) define the two pressures
between which the output signal can generally
change. Both values are adjustable across the en-
tire basic measuring range. The set values always
refer to pressure (in the relevant measuring unit)
and are converted when the measuring unit is
changed.
The assigned signal values for
MA
and
ME
are invari-
able (type label, e.g. 0…10 V or 4…20 mA).
If the parameter
MA
and/or
ME
is
changed, a support point table that is
active at this time will be deleted (cf.
5.3.8 or 5.3.9).
If
MA
is smaller than
ME
this is referred ´to a rising
characteristic curve. The output signal grows as the
pressure increases.
If ME is smaller than MA, this is a decreasing char-
acteristic curve and the output signal decreases
with the falling pressure.
3
0.08 mbar
≙
8 Pa