  Attachment and utilization of the device 

 

 

 

 

 

 

1. Attach the electrodes to 

the cable and device end. 

2. Remove the cover. 

3. Attach the device end to 

the 

right

 side of the body 

under the collarbone. The 

cable end is attached to the 

left

 side of the body on the 

rib cage. 

4. The measurement starts 

automatically when the 

device is attached. Make 

sure the green led is 

flashing. Note: The light is 

easiest to see in a dark 

room. 

 

Detach the device during showers, sauna and swimming. It is not allowed to wear the device in the water! 
Recording continues automatically when you re-attach the device. When you want to end the measurement, 
simply detach the monitor from your body and remove the electrodes. The device will stop recording 
automatically. 

The electrodes are disposable. Change them at least once per day, for example after taking a shower or if they 
become loose during the measurement. The glue or electrode paste can irritate your skin, so clean and dry your 
skin after removing the electrodes. You can also slightly alter the electrode location between days. 

  
Explanations for signal lights during a measurement 

 

 

Green led is flashing in sync with the 

heart rate. 

 

Device is recording heartbeat data. 

 

Green and orange led are flashing 

simultaneously.  

Battery is low, but device is still recording. If you have 

just started the measurement, get in touch with your 

contact person. 

 

 

Red led is flashing or burns constantly.  

Battery is dead or the memory is full. Recording has 

ended. Stop the measurement and get in touch with your 

contact person.  

 

 