3
1.0 PREFACE
1.1
TECHNICAL SPECIFICATION
Fuels:
Wood
Light Fuel Oil Heavy Fuel Oil
Natural Gas
Propane
Coal 4 - User definable fuels
Accuracy:
O
2
- Better than 1% vol. Other gases - Better than 4% of reading
Display:
40 x 8 Matrix Liquid Crystal, electroluminescent backlighting when in
use
Keyboard:
Tactile membrane (integral with display) function keys and cursors
Printer:
External Serial Printer (Optional on all 1000 Series).
Indicators:
LED type for ON (Power), Stand-by, Service, Charge, Low Batt., Fault
Power Supply:
95 - 265V a.c. +10%, 50-60Hz, 30 Watts. Rechargeable battery 2 x 6V
Ampere hours. Typical 8 hr. operation typical.
Sensor type:
O
2
, CO,
Electrochemical Cells
NO (FT1300-1 only)
Electrochemical Cell
SO
2
(FT1200-2A only)
Electrochemical Cell
Flue Gas Temp.:
Type K Thermocouple
Ambient Temp.:
Solid State Sensor
Draft:
Pressure
Transducer
Calibration Check Program:
3 minute zero calibration check on all cells (O
2
check 20.9).
Special ranges are determined before purchase. These ranges only apply where requests have been
made at the time of purchase.
Gas cells
Range
Accuracy
% full scale
Resolution
Oxygen, O
2
0 to 25.0% Vol.
±1%
±0.1% of volume
Carbon Monoxide, CO
0 to 2000ppm
±4%
±1ppm
Nitrogen Monoxide, NO
(FT1300-1 only)
0 to 1000ppm
±4%
±1ppm
Sulfur Dioxide, SO
2
(FT1200-2A)
0 to 2000ppm
±4%
±1ppm
Ambient Temp
-5 to 50°C/+23 to 122°F
Draft
±20” Water Gauge
Carbon Dioxide, CO
2
Calculated
NOx
Calculated (FT1300-1 only)
Efficiency
Calculated
Excess air
Calculated
Loss Calculated
Special ranges:
Carbon Monoxide, CO
0 to 500ppm
±4%
±1ppm
0 to 1000ppm
±4%
±1ppm
0 to 10%
±100ppm
Nitrogen Monoxide, NO
(FT1300-1 only)
0 to 500ppm
±4%
±1ppm
0 to 2000ppm
±4%
±1ppm
0 to 4000ppm
±4%
±1ppm